Cellicon

TM

 Perfusion Solution for Process Development User Guide 

8

Prime System Guidance Step 3

Prime System Guidance Step 4

The 

Next

 button will progress to the next 

step in the manual priming procedure. The 

Back

 button will revisit the previous step. 

When all the steps are complete, select 

Done

. Select 

Exit priming

 to return to the 

main screen at any point.
When 

Done

 or 

Exit priming 

is selected, the 

system returns to flow setpoint mode.

Zero Adjust

Select 

Zero adjust

 from the control region of 

the main screen. A popup containing options 

for zero adjustment will appear.

If connected, the following sensors are 

available for zero adjustment: 
•  Pressure 1
•  Pressure 2
•  Pressure 3
•  Flowmeter 1
•  Flowmeter 2
Select an icon to initiate a zero adjustment 

of that sensor. Adjustment may take a few 

minutes to complete. 
When the adjustment is complete, the button 

will return to its original state. The time and 

date of the last zero adjustment for each 

sensor will be displayed below the button. If 

an adjustment fails, the button will turn red 

and read 

Zero adjustment failed MM/DD/

YY HH:MM.

To initiate a zero adjustment for flowmeter 1, the 

pump must be stopped. A pop-up will appear to 

confirm that the pump is stopped. Flowmeter 2 

does not require that the pump is stopped.
All sensors should be zero adjusted prior to 

starting a run. Only one sensor can be zero 

adjusted at a time. 

Control Modes

The pump can be run in either 

Flow setpoint

 

or 

Speed setpoint

. Choose the mode using 

the toggle switch at the top of the control 

region of the main screen.
When switching from flow control mode 

to speed control mode while the pump is 

running, the controller will default to the 

current pump speed (RPM). Adjust the speed 

using the 

+

/

-

 icons or enter a new speed in 

the field.
When switching from speed control mode to 

flow control mode while the pump is running, 

the controller will default to the last flow 

setpoint (mlpm) entered by the user. Adjust 

the setpoint using the 

+

/

-

 icons or enter a 

new flow target in the field.

Flow Setpoint

Select 

Flow setpoint

 when a perfusion run 

is in progress. Enter the desired flow rate 

and the system will maintain the flow rate 

by adjusting the pump speed (RPM). Speed 

adjustments are based on feedback from 

Flowmeter 1.

Speed Setpoint

Speed setpoint should not be selected 

when a perfusion run is in progress. When 

switching to speed setpoint, a confirmation 

popup will appear. Enter the desired RPM 

and the system will maintain the speed. Use 

speed setpoint for priming.

Summary of Contents for Milipore Cellicon Perfusion Solution for Process Development

Page 1: ...n Solution for Process Development User Guide Contents Introduction 2 System Components 2 Setting Up the System 3 Starting the System 4 User Interface 4 Settings 7 Running the Pump 7 Alerts 9 Settings...

Page 2: ...scribes the installation setup and use of the CelliconTM Perfusion Solution for Process Development This system is for industrial use only and cannot be used in a residential environment System Compon...

Page 3: ...ors on the assembly as follows Press 1 to the feed line Press 2 to the retentate line Press 3 to the perfusate permeate line Connecting the Bioreactor Use a tubing welder to connect The cell retention...

Page 4: ...on Powering up the system launches the perfusion control software The initial screen is the username selection screen After selecting the username the password screen will appear Touch the white box o...

Page 5: ...ton Selecting the Settings button opens the settings screen User Icon and Time Date Display Selecting the user icon allows the user to lock the screen or switch users Lock screen will cause the contro...

Page 6: ...low Setpoint Prime System Zero Adjust Flow Rate Speed Setpoint Pump Start Decrease Flow Rate Increase Flow Rate Status The current values state of the system components and data logging switch are dis...

Page 7: ...ton will read START PUMP If the pump is on the button will read STOP PUMP Stopping the pump will cause a confimation screen to appear Prime System The pump needs to be primed to reduce damage to cells...

Page 8: ...t be stopped A pop up will appear to confirm that the pump is stopped Flowmeter 2 does not require that the pump is stopped All sensors should be zero adjusted prior to starting a run Only one sensor...

Page 9: ...splayed Alarm Acknowledgement Alerts can be removed from the main screen by selecting Acknowledge from the alert bar or in the alert log The alert will reappear on the alert bar if it remains active 1...

Page 10: ...ill make a loud beep when the alert is activated Select Dismiss to return to the main screen then select Acknowledge on the alert bar or within the alert log The data region can still be viewed when a...

Page 11: ...Critical Alarm High Optional 1 Pressure Alarms Pressure Critical Alarm Low Optional 1 Pressure Warning Alarm Low Never 2 Pressure Warning Alarm High Never 2 Pressure Critical Alarm High Optional 1 Al...

Page 12: ...ending is fixed and generated with the date followed by the time when the logfile is created When data logging is switched On a prompt to enter a filename will appear This filename will replace the t...

Page 13: ...error The resolution of the flow signal can be improved by decreasing the full scale of the flowmeter The resolution of the flow signal is always 1 10 000 of the full scale Decreasing the full scale b...

Page 14: ...Refer to Table of Alerts for more information Critical alarms are shown in red color type 1 alert Warning alarms are shown in yellow color type 2 alert Neither critical nor warning alarms will shut d...

Page 15: ...nalog Input PLC Digital Input PLC Analog Output User Accounts Settings The administrator can generate delete and edit user accounts including password protection capabilities and rescrict access permi...

Page 16: ...s have been added to the reactor This will eliminate the possible introduction of cell shear when the system is primed see Priming Preparation Equipment Needs The following user supplied items may be...

Page 17: ...r or gas in the pump head will negatively affect pump performance and decrease the recirculation flow rate MobiusTM 3L Single use Bioreactor When using the MobiusTM single use 3L reactor the inlet to...

Page 18: ...lect Prime system on the main menu screen This will switch the controller into Speed setpoint mode and provide step by step guidance on how to complete the priming 2 Ensure that the perfusate permeate...

Page 19: ...ved on the USB drive the log interval and the log file size Choose a log interval of 60 seconds or greater and a log file size of 65000 lines Insert the USB drive into a computer to view data Open Mic...

Page 20: ...ars light grey and does not read values Connect cables from component to controller Install correct software Component is in an alarm state appears red or yellow Plug in component cables Pump alert Pl...

Page 21: ...d ensure it does not pull air from sparger Check system mode flow control vs speed control No flow through permeate Ensure permeate line is not clamped or pinched Symptom Corrective Action Early filte...

Page 22: ...un Disposing of the Filter The filter is completely single use and should be disposed of after use Follow company procedure for waste disposal Cleaning the Hardware All parts of the system should be w...

Page 23: ...s All other trademarks are the property of their respective owners Detailed information on trademarks is available via publicly accessible resources MK_UG5285EN Rev 1 06 2020 2020 Merck KGaA Darmstadt...