Maintenance
29
Cleaning & Sanitization
Cleaning the inlet strainer
The purpose of the inlet strainer is to prevent large particles from entering the Milli-Q
®
system.
If the inlet strainer becomes clogged, then feed water does not flow freely to the system.
Select the
Care/Cleaning
app and click on
Inlet strainer clean
. Follow the self-guided wizards.
Cl
2
Cleaning
Follow the self guided wizard. Contact your service representative for further information. It is
recommended to perform a Cl
2
cleaning every 90 days.
pH Cleaning
Follow the self guided wizard. Contact your service representative for further information.
Volumetric dispensing calibration
The purpose of this function is to calibrate the volumetric dispensing function. Calibration shall be
perfomed after each replacement of the cartridges or the POD-Paks. Regular calibration will ensure
accuracy of your dispensing.
Note: You will need a 1.0 L graduated cylinder.
Select
Care/Cleaning
app and click on
Volumetric dispensing calibration
. Follow the self-guided
wizards.