Self Help Guide
32
SELF HELP GUIDE
Troubleshooting tips
Turn system off
Press the ON/OFF switch located on the left side
of the unit.
Note:
When the system is on, the power switch is
lit.
&
UNPLUG
Depressurize system
Depressurize system in case of water leak
to temporarily stop production and all water processes.
Go to page 31
Maintenance > Depressurization
.
Identify the source of the leak. Once fixed, to exit this mode, a user will need to go back to the
home screen on which the initial depressurization took place and select
Exit maintenance
for
dispensing to become available again on the Q-POD
®
unit.
Important!
Never open the system, even when the power switch has been switched off. Hazardous
voltage is present and can cause death or serious injury.
It is not sufficient to stop all power to system, the plug must be physically removed from its power
source. Installation and maintenance should only be done by a qualified person. Appropriate
personal protective equipment (PPE) must be worn and safe work practices must be followed.
If the system is not electrically powered for 20 days or longer, then the date is no longer stored on its
circuit board. If this happens, the date and time need to be re-entered, see page 24
Settings>Date
Time & Language
. It takes about 45 minutes to recharge the capacitor.
Inaccurate volumetric dispensing
Volumetric dispensing has been validated to work ± 3% precision. Go to page 29
Maintenance>Cleaning & Sanitization
for more information.