background image

 

Requirements and specifications

38

System specifications

Electrical

The power supply converts mains voltage to 28 VDC. The power supply is compatible worldwide

.

Catalog

Frequency

Max power

Voltage

ZEQ7008T0C

50/60 Hz ± 2 Hz

200 VA

100 – 240 V ± 10%

ZEQ7016T0C

50/60 Hz ± 2 Hz

200 VA

100 – 240 V ± 10%

Environmental

Specific environmental conditions have to be respected to ensure normal operation of the Milli-Q

®

 

water system.

Location

Indoor use only

Ambient operating temperature

Room temperature ranges from 4°C to 40°C

Relative humidity

Room relative humidity conditions of 80% for a 

temperature up to 31°C, decreasing linearly to a 

relative humidity of 50% at 40°C

Altitude

Up to 3000m above sea level

Installation category

Bench / underbench or wall mounted

Pollution degree

2

Means of protection

Class I (PE Connected)

Overvoltage category

II

Communication

  

Each HMI has a large HD capacitive 7″ touch screen (Resolution: 800*480) that allows control and 

monitoring of the system.

 

USB

The HMI has a built-in USB port that offers the possibility to export the system data and/or history. 

The Host interface is compliant with the USB 2.0 High-speed standard.

 

 

USB keys only work when formatted FAT32. NTFS format is not compatible.

Ethernet

When connected through an Ethernet protocol, the display interface can be accessed remotely using 

internet web browsers.

For best browsing performance, Google Chrome™ browser is recommended.

Summary of Contents for Milli-Q EQ 7008

Page 1: ...SigmaAldrich com User Manual Milli Q EQ 7008 7016 Ultrapure Water Systems...

Page 2: ...INFORMATION MENU 16 System 16 History 16 Flow schematic 17 SETTINGS MENU 18 Manager access password 19 System configuration 20 HMI configuration 21 Accessories 22 Alarm configuration 23 Connectivity...

Page 3: ...7016 system is intended to produce ultrapure type 1 and pure water from a tap water source primarily for use in research and quality control in a variety of laboratories worldwide The product is desig...

Page 4: ...ine interface display and Q POD unit Point of Dispense of ultrapure water 1 Front view of benchtop located Milli Q EQ 7016 with Q POD unit and HMI 4 Rear view of a Milli Q EQ 7016 system 2 HMI display...

Page 5: ...le OPTION 4 System feed pressure sensor 22 Distribution inlet solenoid valve 5 IPAK Gard pretreatment pack 23 Recirculation loop 6 Pressure regulator 24 Distribution pump 7 RO recovery loop 25 TOC ind...

Page 6: ...d pressure are monitored Tank is configured with a 2 or 5 meters connectors to the main production unit depending on the laboratory setting The RO water stored in the tank is purified one step further...

Page 7: ...2 The system is powered on STEADY The water quality fits the requirements water quality values are below alarm setpoints BREATHING The system is in recirculation mode A water quality alarm is raised A...

Page 8: ...ulation icon this also refreshes the water quality parameters Product resistivity TOC Temperature There are several options to dispense ultrapure water Manual Dispensing To start dispensing in low flo...

Page 9: ...is reached Water Quality values When the dispensing is finished the resistivity and temperature are instantly displayed on the left side of the HMI Ninety seconds later a new TOC value is displayed Th...

Page 10: ...ction to help the user troubleshoot If the issue is not resolved and depending upon the alert and alarm either select the Contact tab to find the technical support hotline number for further assistanc...

Page 11: ...Dispensing functions Alerts Alarms when generated by the system Water level in tank Recirculation start icon MAIN MENUS Menu Settings Information Maintenance Consumables Login Go to Consumables Menu...

Page 12: ...n screens Pops up screens When a text box is selected the keypad will automatically appear To remove the keypad click on or on any zone outside the keypad Note When in a menu or app level screen the s...

Page 13: ...g agents that can be used include KLERCIDE Isopropanol 70 or equivalent composition SPOR KLENZ Ready to use or equivalent composition Be sure to follow the safety precautions listed with the recommend...

Page 14: ...ble the ordering information and the consumable s ID Certificates of quality are available online www mymilliqconsumables com All cartridges have been designed to be easily replaced by a user A shortc...

Page 15: ...installation date The IPAK Quanta and Meta polishing cartridges IPAK Gard pretreatment pack and application POD Paks information is automatically registered during their installation The tank vent fi...

Page 16: ...spensing events report Each row contains a record of a single dispensing It is an archive of all the dispense reports To export data or preview a graph 1 Select a time period of interest with a start...

Page 17: ...quality consumables installed dates Active alerts alarms auto test results Diagnostics Contacts To edit contact information login as the system manager go to page 19 Settings Manager access The user...

Page 18: ...pacity Read only Read only Tank refill setpoint Read only ech2 o ASM lamp Read only Recirculation duration Read only Read only Safety flow stop Read only Custom volume Connectivity Local network Read...

Page 19: ...k HMI configuration Brightness Buzzer Read only Date Time Lan guage Time zone Date Time Read only Language Read only Password System manager Units Resistivity Conductivity Read only Temperature Read o...

Page 20: ...ot going to be used for more than 3 continuous days then the Lab Close mode can be activated in order to save energy Before starting this mode the tank is manually emptied During this mode periodic re...

Page 21: ...lush check time By default at 1 a m the system rinses the RO cartridge s for 3 minutes unless the system has operated more than 1 hour in the past 24 hours This is done to maintain water purity and fr...

Page 22: ...g the volumetric dispensing mode Washer Distribution kit Contains a pump and pressure switch Used to pump water directly from the storage tank to an application such as a dishwasher It does not interf...

Page 23: ...365 1 365 Tank empty 0 0 50 Ultrapure resistivity M cm 25 C 16 5 1 18 2 Ultrapure TOC ppb 500 0 1 999 IPAK polishing cartridges days 365 0 365 Millipak 0 22 m filter days 182 30 182 Millipak Gold 0 22...

Page 24: ...ndar date representations are shown in the ISO 8601 format 2018 11 06 YYYY MM DD YYYY indicates a four digit year MM indicates a two digit month of the year 01 through 12 DD indicates a two digit day...

Page 25: ...d to 25 C The values are normalised The system firmware eliminates small fluctuations of temperature compensated resistivity or conductivity due to the fact that these two parameters are not measured...

Page 26: ...up 3 A shortcut from a consumable pop up IPAK Quanta and IPAK Meta polishing cartridges replacement 1 Select Install consumables app and click on Install IPAK Meta Quanta 2 Follow the self guided wiza...

Page 27: ...ol should face inwards Note 2 If the sound has been activated it is normal to hear a buzzer when changing the consumables This is temporary and is only present when the IPAK cartridges or POD Pak are...

Page 28: ...card that is used to register its installation Important Due to a UV lamp located inside the storage tank never remove the circular lid You can register the vent filter details by selecting the Insta...

Page 29: ...representative for further information It is recommended to perform a Cl2 cleaning every 90 days pH Cleaning Follow the self guided wizard Contact your service representative for further information V...

Page 30: ...ified person Appropriate personal protective equipment PPE must be worn and safe work practices must be followed Cleaning the external surfaces For cleaning and disinfecting the external surfaces of t...

Page 31: ...s temporarily stops production and all water processes To go back to tank full tank filling swipe down from the top of the touchscreen and press Distribution maintenance The distribution maintenance i...

Page 32: ...m even when the power switch has been switched off Hazardous voltage is present and can cause death or serious injury It is not sufficient to stop all power to system the plug must be physically remov...

Page 33: ...to check the status of the pure production section Exit the maintenance or rinsing mode in case it is activated Solve any alarms that could be stopping the production process At the end make sure the...

Page 34: ...set Remove BOTH old cartridges Then insert new cartridges Check that the e Sure tag is correctly working A pre scan can be initiated by entering into the IPAK Meta Quanta installation wizard If the e...

Page 35: ...1 Alarm with number of active alarm s Alert Home Menu Storage tank System manager logged in Mass storage USB connected Ethernet LAN status connected Slider ON Slider OFF Back Calendar entry Password...

Page 36: ...th size 0 22 m with Millipak filters Bacteria 0 01 CFU mL with Millipak and Biopak filters 0 005 CFU mL with Millipak Gold installed and used in a laminar flow hood Pyrogens endotoxins 0 001 EU mL wit...

Page 37: ...nese National Standard GB T 6682 Grade 1 water Chinese National Standard GB T 33087 Ultrapure water JIS K 0557 A4 water Clinical and Laboratory Standards Institute CLSI Clinical Laboratory Reagent Wat...

Page 38: ...sing linearly to a relative humidity of 50 at 40 C Altitude Up to 3000m above sea level Installation category Bench underbench or wall mounted Pollution degree 2 Means of protection Class I PE Connect...

Page 39: ...t 15 Radio frequency devices Sub part C Intentional Radiators Limits and Methods of measure ment of radio disturbance System software System software included in this product contains copyrighted soft...

Page 40: ...ds upon the water system type EQ 7008 or EQ 7016 and feed water quality an average of 3 1 L of feed water is needed to produce 1 L of type 1 purified water Dimensions and weights Water purification sy...

Page 41: ...lb 0 42 kg 0 9 lb Water storage 25 L 6 7 kg 14 8 lb 8 5 kg 18 7 lb 31 7 kg 69 9 lb Water storage 50 L 7 6 kg 16 8 lb 10 6 kg 23 4 lb 57 6 kg 127 lb Water storage 100L 10 9 kg 24 0 lb 12 8 kg 28 2 lb...

Page 42: ...esentative or organization Ordering information System components Description Catalog Number Wall mounting Kit for Q POD includes Q POD unit WMEQ0RKT System mounting kit for Q POD includes Q POD unit...

Page 43: ...s consumable kit IPAKKITA1 IPAK Gard pretreatment pack IPAKGARA1 IPAK Gard pretreatment pack with polyphosphate IPAKGAR1H UV Lamp ZEQ7UVLP0 Millipak 0 22 m filter MPGP002A1 Millipak Gold 0 22 m steril...

Page 44: ...oducts are non contractual Trademarks New trademark names The initial M Millipore Milli Q Q POD ech2o IPAK Gard IPAK Meta IPAK Quanta Jetpore IQnano Millipak Biopak LC Pak EDS Pak and VOC Pak are trad...

Page 45: ...e of electrical power that is earth grounded Attention Before the system is serviced unplug the electrical power cord The Milli Q system must be powered OFF before plugging in or removing any componen...

Reviews: