Installation Instructions

 1.  Place the unit on a sturdy and level surface in a 

safe dry place away from laboratory traffic.

 2.  Ensure that the power switch is OFF, and then plug 

one end of the three-pronged power cord into a 

grounded three-prong AC outlet with appropriate 

voltage (100 V to 240 V, as indicated on the rating 

sticker near the AC cord on the back of the unit) 

and plug the other end into the main power socket. 

Power Switch

 3.  Connect the DC output jacks from the 

electrophoresis unit; insert the red lead (+) into  

the red output jack, and the black lead (-) into  

the blue output jack.

 4.  Power ON the unit by pressing the ON / Off switch  

on the back.

Operating Instructions

Control Interface

LED Indicator 

Indicates running 

conditions

Button Functions

Key 

Descriptions

Icon

Function

Start 

Stop

Press to activate or stop the unit.

Pause

Press to temporarily interrupt power to an 

operation in progress; resume power after 

pausing without resetting the timer.

Page 

Toggle

Press to go back to previous page.

Enter

1. Press to enter the next page. 

2. Press to enable numeric set-up.

Left 

Arrow

Press to move cursor left forward between 

parameters.

Right 

Arrow

Press to move cursor right forward 

between parameters.

Up  

Arrow

1. Press to move cursor up between 

parameters. 

2. Press to increase numeric values.

Down 

Arrow 

1. Press to move cursor down between 

parameters. 

2. Press to decrease numeric values.

Plugs

Summary of Contents for Millipore mA700

Page 1: ...er Constant voltages constant currents and constant power Four terminal outlets LCD display Timer with alarm function Advanced safety devices Stackable Wide applications for DNA RNA and protein electrophoresis User Guide mA700 Essential Power Supply mA700 Introduction 1 Equipment Operation 2 Symbols Used on Device Regulatory Warnings Safety Precautions Environmental Operating Conditions Avoiding E...

Page 2: ...carefully before operating this equipment Do not alter the equipment Failure to follow these directions could result in personal and or laboratory hazards as well as invalidate equipment warranty Use a properly grounded electrical outlet with correct voltage and current handling capacity Disconnect from power supply before maintenance and servicing Refer servicing to qualified personnel Never use ...

Page 3: ...erate the power supplies in high humidity environments 95 or where condensation may occur 7 To avoid condensation after operating the power supply in a cold room wrap the unit in a plastic bag and allow at least 2 hours for the unit to equilibrate to room temperature before removing the bag and operating the unit 8 Prior to applying any cleaning or decontamination methods other than manufacturer s...

Page 4: ...jack 4 Power ON the unit by pressing the ON Off switch on the back Operating Instructions Control Interface LED Indicator Indicates running conditions Button Functions Key Descriptions Icon Function Start Stop Press to activate or stop the unit Pause Press to temporarily interrupt power to an operation in progress resume power after pausing without resetting the timer Page Toggle Press to go back ...

Page 5: ...ower supply will constantly operate until the user manually ends the run Constant Mode Operation 1 Press the Start Stop key to start operation The LED is lit and the screen will show real time parameter values When parameter reaches the set value its color will turn orange 2 Press the Toggle key to switch the display between real time value and set value page Note When a run is completed operation...

Page 6: ...Press the Start Stop key to start operation The LED is lit and the screen will show real time parameter values When parameters reach the set value the color will turn orange 2 Press the Toggle key to switch the display between real time value and set value page Note When a run is completed operation stops with an alarm and COMPLETE is displayed on the screen 3 Press the Start Stop key to terminate...

Page 7: ...blown Replace the fuse Fuse broken repeatedly Hardware failure Contact our Technical Service Department Operation stops and the screen displays PLC ERROR Communication wires on circuit board have loosened or broke We recommend that you send the machine back to your local distributor or our Technical Service Department for maintenance Operation stops with alarm The screen displays NO LOAD Electroph...

Page 8: ... with alarm Display shows LEAKAGE Ground leak detected during run Check the electrophoresis system for improper grounding Restart the power supply by turning the Power switch off and on Operation stops with alarm Display shows OVER TEMPERATURE Power supply is overheating Turn off power supply Check for sufficient airflow around the power supply fan After cooling down restart the power supply by tu...

Page 9: ...ower Supply with cord for France Germany and locations using CEE 7 7 YP22 plugs MA400 EU mA400 Power Supply with cord for the UK Ireland and locations using type G YP61 plugs MA400 UK mA400 Power Supply with cord for Japan and locations using YP18 plugs MA400 NI mA400 Power Supply with cord for China and locations using Type 1 YP03 plugs MA400 ZH Support Notice We provide information and advice to...