Service Manual Version 18

7

.5.1

 

 

AUGUST 3, 2021   SQA-iO Service Manual                     

4

 

 

      

 

Technology 

Motility Channel 

Light disturbances caused by moving sperm cells are converted into 

analog

 

signals.  

Proprietary algorithms interpret the signals and report motility. 

 

Concentration Channel 

Based on spectrophotometry analysis of the semen samples and application of 
proprietary algorithms.   
 

 

 

1.

 

The capillary is inserted into the measurement compartment. 

2.

 

Sample concentration is evaluated in the "tall" 10 mm chamber of the capillary.

 

3.

 

Motility is detected in the "thin" 0.3 mm section of the capillary.

 

4.

 

This information is then digitized and routed to the microprocessor that applies 
algorithms to extract the required clinical semen parameters. 

 

Analog to 

Digital 

Converter

Semen 

parameters 

extraction 

algorithms

Self-Test 

and QC

KeyPad

M

ic

ro

co

nt

ro

ll

er

Operational 

Monitor

#1

#2

#3

#4

Concentration 

detector

Motility 

detector

Concentration 

LED

Motility 

LED

 
 

SQA-iO Device 

Overview and 

System 

Components  

 
 
 
 
 
 
 
 
 
 
 

 

 

Measurement 

Compartment 

 
 
 
 
 
 
 
 
 
 

 
SECTION II: System Specifications and Requirements   

 

 

Dimensions: 8 X 9.5 X 10.5 cm  

 

Weight: 0.350 Kg  

 

Power supply: 5V DC (USB) 

 

Noise level: 0 [dBA] 

 

Device power consumption: 1.7 [BTU/hour] = 0.5 [Watts] 
 

Minimum requirements

:

 

 

PC: Intel Core i5 M520 2.4GHz or equivalent  

 

RAM: 4GB 

 

Monitor Screen: Color, Wide screen 

 minimum resolution 1024 x 768   

 

Operating system compatibility: Windows 7 Professional x32 or above  

 

Communication Ports: one USB port  

 

Internet Access: 5mb per second 

Front Panel:  Measurement Compartment  (Capillary insertion for testing)

 

 

Capillary

 

chamber

 

Summary of Contents for SQA-iO

Page 1: ...Service Manual Version 187 5 1 AUGUST 3 2021 SQA iO Service Manual 1 Service Manual Version 187 5 1 Catalog IO ML 01678 00 Rev August 2021...

Page 2: ...t 4 USB Port 5 Testing Capillary 5 Maintenance Schedule 5 Testing and Operting Requirements 6 Operating Temperature Humidity Altitude 6 SECTION III Quality Control Internal QC 6 Printing the Service R...

Page 3: ...ple is withdrawn into an SQA capillary disposable delivery system inserted into the SQA iO where key parameters are collected and sample test results are processed utilizing proprietary technology and...

Page 4: ...s to extract the required clinical semen parameters Analog to Digital Converter Semen parameters extraction algorithms Self Test and QC KeyPad Microcontroller Operational Monitor 1 2 3 4 Concentration...

Page 5: ...ng Capillary Sample delivery system Normal sample Requires approximately 0 5cc of semen Short samples Requires approximately 20 l Single use design for testing semen in a biologically safe manner Moti...

Page 6: ...sing linearly to 50 humidity at 38 C Operates in a wide range of ambient temperatures 15 38 C however the system is calibrated to measure semen samples at room temperature 20 25 C 68 77 F Intended for...

Page 7: ...nge Instructions for printing the SQA iO SERVICE parameters to prepare for technical support If a Self Test failure occurs the related status icon in the SQA iO HOME SCREEN will turn red Click on the...

Page 8: ...f the optics you will feel a shelf in the back top section of the chamber Fig 2 and 3 CLEANING STEP 2 1 Use a Fibrous material cleaning paddle Fig 4 supplied in your TEST KIT Moisten with only ONE dro...

Page 9: ...icon indicating a failure Observe the normal range value 1 Clean the test chamber using ONLY MES cleaning kit 2 Remove and reconnect USB cable 3 Reboot the device If problem persists refer to MES cust...

Page 10: ...de passed the self test Amplitude failed the self test Suggestion Clean the device ZERO LEVEL 500 525 Zero Level passed the self test Zero Level failed the self test Suggestion Clean the SQA iO Mainta...

Page 11: ...ting the sample collection container WARNING Do not shake or use a pipette to mix the sample otherwise air bubbles will form and test results will be inaccurate Figure 1 5 Carefully check that the liq...

Page 12: ...ary both top and bottom Figure 4 with a delicate wipe Kimwipes etc It is important to remove all semen from the exterior of the capillary in order to keep the SQA iO clean Visually confirm that the ca...

Page 13: ...section while keeping the tip of the capillary well below the sample level and well below the level of any bubbles covering the liquid 3 Withdraw the capillary tip from the semen sample and visually i...