Service Manual Version 18

7

.5.1

 

 

AUGUST 3, 2021   SQA-iO Service Manual                     

7

 

 

      

 
 

Controls Run 

Prior to   

Sample  

Testing 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Printing the 

Service 

Parameter 

Report 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Autocalibration  verification

:  Reference  parameters  of  the  concentration  and 

motility channels are measured again (without a testing capillary).  

 

System  noise:

  Measures  the  electronic  noise  level  of  the  system  to  insure 

effective measurement of electronic signals. The system automatically adjusts the 
noise level threshold to insure accurate readings.  

 

Electronic spikes:

 Checks for measurement points that are out of range. 

 
 

Instructions for printing the SQA-iO SERVICE parameters to prepare for 
technical support: 

If a Self-Test failure occurs, the related status icon in the SQA-iO 

HOME SCREEN

 will turn 

red. Click on the 

RED

 icon to view instructions for how to resolve the problem: 

 

 

 

 
 
 

SECTION IV: TROUBLESHOOTING   

The SQA-iO troubleshooting guide is focused on app access and function.   The 
SQA-iO device cannot be opened for service so it is important to follow the cleaning 
and use instructions for optimal and continued success using the device.   

Please refer to the appendix section for a TROUBLESHOOTING GUIDE.  

Support is available through your local distributor online.  Please contact them 
directly for questions concerning the device that are not outlined in this service 
manual.     

 

 
 
 
 
 
 
 
 
 

Summary of Contents for SQA-iO

Page 1: ...Service Manual Version 187 5 1 AUGUST 3 2021 SQA iO Service Manual 1 Service Manual Version 187 5 1 Catalog IO ML 01678 00 Rev August 2021...

Page 2: ...t 4 USB Port 5 Testing Capillary 5 Maintenance Schedule 5 Testing and Operting Requirements 6 Operating Temperature Humidity Altitude 6 SECTION III Quality Control Internal QC 6 Printing the Service R...

Page 3: ...ple is withdrawn into an SQA capillary disposable delivery system inserted into the SQA iO where key parameters are collected and sample test results are processed utilizing proprietary technology and...

Page 4: ...s to extract the required clinical semen parameters Analog to Digital Converter Semen parameters extraction algorithms Self Test and QC KeyPad Microcontroller Operational Monitor 1 2 3 4 Concentration...

Page 5: ...ng Capillary Sample delivery system Normal sample Requires approximately 0 5cc of semen Short samples Requires approximately 20 l Single use design for testing semen in a biologically safe manner Moti...

Page 6: ...sing linearly to 50 humidity at 38 C Operates in a wide range of ambient temperatures 15 38 C however the system is calibrated to measure semen samples at room temperature 20 25 C 68 77 F Intended for...

Page 7: ...nge Instructions for printing the SQA iO SERVICE parameters to prepare for technical support If a Self Test failure occurs the related status icon in the SQA iO HOME SCREEN will turn red Click on the...

Page 8: ...f the optics you will feel a shelf in the back top section of the chamber Fig 2 and 3 CLEANING STEP 2 1 Use a Fibrous material cleaning paddle Fig 4 supplied in your TEST KIT Moisten with only ONE dro...

Page 9: ...icon indicating a failure Observe the normal range value 1 Clean the test chamber using ONLY MES cleaning kit 2 Remove and reconnect USB cable 3 Reboot the device If problem persists refer to MES cust...

Page 10: ...de passed the self test Amplitude failed the self test Suggestion Clean the device ZERO LEVEL 500 525 Zero Level passed the self test Zero Level failed the self test Suggestion Clean the SQA iO Mainta...

Page 11: ...ting the sample collection container WARNING Do not shake or use a pipette to mix the sample otherwise air bubbles will form and test results will be inaccurate Figure 1 5 Carefully check that the liq...

Page 12: ...ary both top and bottom Figure 4 with a delicate wipe Kimwipes etc It is important to remove all semen from the exterior of the capillary in order to keep the SQA iO clean Visually confirm that the ca...

Page 13: ...section while keeping the tip of the capillary well below the sample level and well below the level of any bubbles covering the liquid 3 Withdraw the capillary tip from the semen sample and visually i...

Reviews: