Service Manual Version 18
7
.5.1
AUGUST 3, 2021 SQA-iO Service Manual
7
Controls Run
Prior to
Sample
Testing
Printing the
Service
Parameter
Report
•
Autocalibration verification
: Reference parameters of the concentration and
motility channels are measured again (without a testing capillary).
•
System noise:
Measures the electronic noise level of the system to insure
effective measurement of electronic signals. The system automatically adjusts the
noise level threshold to insure accurate readings.
•
Electronic spikes:
Checks for measurement points that are out of range.
Instructions for printing the SQA-iO SERVICE parameters to prepare for
technical support:
If a Self-Test failure occurs, the related status icon in the SQA-iO
HOME SCREEN
will turn
red. Click on the
RED
icon to view instructions for how to resolve the problem:
SECTION IV: TROUBLESHOOTING
The SQA-iO troubleshooting guide is focused on app access and function. The
SQA-iO device cannot be opened for service so it is important to follow the cleaning
and use instructions for optimal and continued success using the device.
Please refer to the appendix section for a TROUBLESHOOTING GUIDE.
Support is available through your local distributor online. Please contact them
directly for questions concerning the device that are not outlined in this service
manual.