Revision 21_Sep_2015
45
User Guide Version 2.60 I-button WHO 5th
APPENDIX 14: SQA-V SERVICE SUPPORT REPORT
SQA-V SERVICE SUPPORT
Parameter Report
Device number:
SQA-V Software Version:
Date:
Instruct the user to run a SERVICE report. For version 2.60 from the MAIN MENU select:
SERVICE > PRINT
SELF TEST DATA
.
Calibration parameters:
Fill-in the USER REPORT column with the calibration parameters found in the INTERNAL DATA SECTION of the
SERVICE DATA REPORT run on the “defective” SQA-V. Contact MES for the initial calibration parameters. These
parameters should not have changed.
Parameter
Service
Report
Item #
User Report
MES
Report
Comments
CONTR.REF1 #1
OD AMPLIF.
#13
MSC AMPLIF
#8
OD VALUE
#15
OD CORR
#16
LB OD AMP
#18
CONTR. Z.L*
#11
*CONTR. Z.L. can be adjusted in the field by a MES trained service technician.
Algorithm parameters:
Fill-in the User Report values for the following algorithm parameters found in the SERVICE DATA REPORT.
The SQA-V algorithm settings are defined and should not have changed.
Parameter
Service
Report
Item #
User
Report
MES
Settings
Comments
MIN.SP.HEIGHT #2
5
MIN.SP.WIDTH #9
10
MAX.SP.WIDTH #3
150
NOISE THRESH
#10
6
SMI THRESH
#4
28