24
3.
90XL Pressure Calibration Verification
Positive or negative gauge pressure may be checked against a mercury
manometer or other N.I.S.T. traceable standard.
To verify calibration of the pressure sensor module:
1.
The 90XL “Ref” port must be open to the atmosphere.
2.
Connect the 90XL pressure sensor module to the display module.
3.
With the “Gauge” port open to atmospheric pressure, the display should
read “0 mmHg”.
NOTE:
If the display does not read zero, refer to the “Calibration and Maintenance”
section of this User’s Guide.
4.
Connect the 90XL and the reference standard using the stop cock and
monitoring lines as shown below. Check for leaks.
5.
Apply positive or negative pressure to both the 90XL pressure sensor
module and the reference standard to verify accuracy.
NIST
Traceable
Pressure
Standard or
Mercury
Manometer