29
2.
Mid-Range (Standard) Conductivity Calibration
PROCEDURE
DISPLAY
· 14.0 or 13.4 mS Reference Solution
· 90XL Sample Kit
· Make sure a condo/temp sensor module is attached.
After “CALIBRATION” has been entered,
Select either 14.00 or 13.4 mS STANDARD from the
next menu list. (14.00 mS is chosen for this example.)
Step 1
- Empty any liquids from the sensor module.
The value of conductivity read internally by the 90XL
must be less than 1 mS to proceed. The factory
calibration value is displayed in this step also.
Step 2
- Rinse the sensor module by drawing
reference solution up through it. The value read
internally must be within 0.2 mS of the reference value
to proceed to the next step. If the factory value is off
by more than 0.2 mS, calibration will not be
performed. Press ABORT and contact Mesa Labs.
Step 3 -
Empty reference solution from the sensor
module, and discard. The 90XL must detect a value of
less than 1 mS to proceed.
Step 4 -
Draw fresh reference solution slowly through
the sensor module. While reference solution is
flowing, the 90XL must detect 4 stable conductivity
values within 0.2 mS of the reference to proceed to
the next step.
NOTE:
Draw slowly so that up to 10 seconds of flow
can be maintained.