– 58 –
Power indicator is
not lit and nebulizer
is not nebulizing.
Batteries installed bac
-
kwards.
Re-install the batteries in
the correct orientation and
restart the power.
(See page 7~8)
Low battery power.
Replace with new batteries
and restart the power.
(See page 7~8)
Incorrect connection of AC
adapter to nebulizer.
Re-connect in the correct
manner and restart the
power. (See page 8)
Power indicator is lit
and nebulizer is not
nebulizing.
Low-power indicator is lit
constantly, insufficient batte
-
ry power, or battery
has run out.
Replace with new batteries
and restart the power.
(See page 7~8)
Rupture of mesh of medica
-
tion chamber.
Replace with a new medica
-
tion chamber and then put in
the medication. (See page 16)
Electrodes on medication
chamber are clogged with
medication or water.
Clear the electrodes of
clogged medication or water
and restart the power.
(See page 14)
Electrodes on nebulizer and
medication chamber are
stained.
Remove the stains and
restart the power. (See
page 14)
Mesh of medication chamber
is severely clogged.
If it still cannot be used after
cleaning, please replace with
a new medication chamber.
(See page 13~16)
Nebulizer shuts off
in usage.
Medication chamber is
loosened and not completely
attached.
Re-attach the medication
chamber correctly and re
-
start the power. (See page 6)
Connection of AC adapter to
nebulizer is loosened.
Re-connect in the correct
manner and restart the
power. (See page 8)
Medication has run out.
Put in the medication in the
medication chamber.
(See page 9~10)
No contact between medica
-
tion and mesh for more than
15 seconds.
Adjust the nebulizer’s angle
so the medication can come
in contact. (See page 11)