GT-521S Manual Rev D
Page 25
7.4.
Comma Separated Value (CSV)
CSV Header (Note 1):
Time,Size1,Count1(CF),Size2,Count2(CF),AT(C),RH(%),Location,Seconds, Status
If differential Mode:
Time,Size1,Count1(dCF),Size2,Count2(dCF),AT(C),RH(%),Location,Seconds, Status
CSV Example Record:
2017-03-23 09:21:29,00.3,00084140,00.5,00022,033,001,0060,000,*03414
Note: If COUNT is set to NORMAL data will be reported as cumulative counts. If COUNT is set
to DIFFERENCE data will be reported as differential counts regardless of setting when data was
collected.
CSV Fields
Field
Parameter
Example Value
Notes
1
Date and Time
2017-03-23 09:21:29
2
Channel 1 Size
00.3
3
Channel 1 Count (TC, /L, CF, M3)
00084140
Note 2
4
Channel 2 Size
00.5
5
Channel 2 Count (TC, /L, CF, M3)
00008680
Note 2
6
Temperature (C,F)
+022
Note 2 & Note 3
7
RH (%)
033
Note 3
8
Location
001
9
Sample Time (0-9999 seconds)
0060
10
Status Bits (see below)
000
Note 5
11
Check sum
*03414
Status Bits
Notes (for table above):
Bit
Value
Condition
0
OK (no alarms or errors)
1. CSV header included for multiple record
transfers like All Data (2) or New Data (3) or
last n data (4 n).
0
1
Count alarm size 1
2. Units determined by product setting.
1
2
Count alarm size 2
3. Temperature and RH will be null (,,) if
Temp/RH probe is not attached
2
4
Not used
4. Alarm sizes will be null (,,) if alarms are
disabled.
3
8
Not used
5. Status bits combinations are possible. For
example, 17 (00010001B) = Low Battery and
Size 1 alarm.
4
16
Low battery
5
32
Sensor error
6
64
Not used
7
128
Not used
7.5.
Printer Format