Accessories and Options
MicroVol-1100 User manual 1.6
To install or remove the adaptor, first separate the 47mm filter holder and impactor
assembly from the inlet tube. Disassemble the impactor assembly. Refer to Figure 1
Section 1.4.2 for details of how the parts a re assembled.
With the filter holder, connecting tube and impactor plate removed, carefully insert
(or remove) the PM
2.5
nozzle adaptor in the nozzle body.
The nozzle body with the PM nozzle adaptor installed.
2.5
Figure 32 PM Nozzle components and assembly
2.5
Carefully reassemble the impactor assembly by inserting the greased impactor plate
(greased side towards nozzle), connecting tube, and filter holder with filter element
installed. Finally, re-attach the assembly to the inlet tube.