19. 5-year inspection
In accordance with the German Waste Management Regulations, amalgam
separators must be examined at intervals of no longer than 5 years.
Conduct the annual inspection, as described in chapter 18.
Check whether the amalgam separator is correctly mounted and
connected.
Rinse the suction hose and the spittoon bowl with at least one
liter of clean water and the specifically designated disinfectant
GREEN
&
CLEAN M2.
43
Visual control of the centrifuge
Wear protective gloves!
Switch the main switch off!
Remove the separator from the wall mounting bracket.
43.1
Remove the four screws on top of the device using a Torx 20
43.2
Remove the lid of the separator (use the screwdriver at the
backside).
Ö
Do not remove the inlet sealing of the separator!
43.3
Detach the separation rotor by removing the plug at the center of
the rotor using pointed pliers, loosening the internal hex nut and
lifting up the rotor. (ATTENTION: left handed threads!)
43.4
Pull off the pump hose at the back of the separator at the upper
connection
43.5
Lift up the centrifuge inlet
43.6
Lift up the centrifuge chambers
Carry out a visual check of the centrifuge chambers for contamination;
centrifuge chambers which show excessive signs of deposits or
sedimentation of solid material must be replaced!
Finally, reassemble in the reverse order. It is important that all screws are
fixed and screwed in as far as they go.
ATTENTION! Make sure the four pins
43.7
are inserted in
the holes on the bottom of centrifugal chamber. Also, the
square of the separation rotor
43.8
must click-fit into the
upper side of the centrifuge
43.9
.
i
The 5-year inspections must always be recorded in the
Equipment Logbook!
5-year inspection
Practice personnel, technicians