13. Description of the external display
15
See illustration
15.1
Control light 1: ready for operation
Ö
green illuminated:
unit is switched on
15.2
Control light 2: centrifuge error
Ö
red flashing and audible buzzer signal: Error!
see illustration
16
error messages
15.3
Control light 3: container full
Ö
yellow illuminated and audible buzzer signal which can be
turned off by pressing RESET: Collection container is 95% full
It is recommended that the collection container be replaced;
however, it is possible to continue operation until the container
is 100% full. As a reminder, the control lamp remains illuminated
and every time the unit is switched on, the buzzer signal is
activated.
Ö
yellow illuminated and audible buzzer signal which cannot be
turned off by pressing RESET: Collection container is 100% full
The collection container must be replaced immediately! Further
operation is not possible, as the suction valve remains closed.
15.4
Alarm RESET button
By pressing the RESET button, the audible 95% full buzzer warning
can be switched off.
16
Error messages
16.1
See illustration
Signal 1
15.1
green illuminated, signal 2
15.2
flashing red, audible
buzzer signal:
Ö
Error in the electronic system!
The buzzer is turned off by pressing the RESET button. Switch the main
button off immediately and after a short break (approx. 5 seconds) on
again. If the error signal remains please contact your service technician!
16.2
See illustration
Signal 2
15.2
flashing red, audible buzzer signal:
Ö
Collection container has not been inserted properly!
The buzzer cannot be turned off, and no other signal illuminates. Please
refer to chapter 14 for the proper replacement of the collection container.
Description of the external display
Practice personnel, technicians