Statement of Intended use
The medical UPMs are intended to protect medical, non-medical
computer equipment and medical devices that need battery backup,
surge protection, voltage regulation and line noise filtering in and around
patient care areas. Safe and continuous operation of the UPM depends
partially on the care taken by users. Please observe the following
precautions. Not following these could result in warranty being voided.
NOTE:
• The UPM is intended for stationary use by itself. If the UPM is
integrated on to a mobile cart, it must be secured to the cart and
tested in the application by the end user.
• This UPM is intended for patient vicinity. AMETEK Powervar
products are not designed for use in any application intended to
support or sustain life.
• Do not use this UPM for life support applications in which a
malfunction or failure of the UPM system could cause failure or
significantly alter the performance of a life-support device.
•
Do not use this UPM near or around flammable gases. Do not use
this UPM within oxygen-enriched atmospheres.
• Do not disassemble the UPM.
• UPM is CLASS 1 equipment.
• Do not attempt to power the UPM from any receptacle except a
properly grounded receptacle that matches the input plug provided
with the UPM.
• Do not place the UPM near water or in environments of excessive
humidity.
• Do not allow liquid or any foreign object to get inside the UPM.
• Do not block air vents on the side of the UPM. Keep a minimum of
3 inches on all sides.
• Do not plug appliances such as hair dryers, fans, heaters, etc. into
the UPM.
• Do not place the UPM under direct sunshine or close to heat
emitting sources (excessively warm temperatures will shorten
battery life).
• This UPM is intended for installation in a temperature controlled,
indoor area free of conductive contaminants.
• The AC power source for the UPM should be conveniently near the
UPM and easily accessible – avoid extension cords or temporary
power strips to power the UPM.