43
LP-80
Table 3 Troubleshooting the LP-80 (continued)
Problem
Possible Solutions
Error message
Firmware is corrupted!
See Manual for options.
This screen means that the firmware was not loaded properly or an
update is needed.
Check
if a new firmware version is
available.
NOTE: Saved data records will be lost when new firmware is installed.
Download any data before reloading the new firmware.
If a newer firmware version is not available, contact
.
Error message
Cannot change settings
while logging!
Configuration settings are locked during active reading or logging
mode to prevent inconsistent data from being saved.
Navigate to the PAR or Log tab and exit the active reading. Make the
desired changes in the Configuration tab and begin a new reading.
Error message
One mode can be run at
a time. Press ENTER to
stop logging and begin
reading.
Only one mode (either PAR or Log) may be running at a time.
Press
ENTER to stop the currently running mode and to continue
with the new mode, or press
MENU
to proceed to the next tab
without changing the mode.
Error message
No summary data.
The view menu only allows you to view summary records. I fyou have
not saved any records or if you only have saved above and below
records, this message will appear when trying to view or download
data. To make this message go away, save a summary recordon the
PAR or Log menu by pressing
ENTER
.
Warning message
By continuing, you
will lose any custom
information. Press ENTER
to continue or ESC to
cancel.
This message will appear when overriding custom location
information.
Press
ENTER
to continue with new information or press
ESC
to
preserve old customer information.
4.5 CUSTOMER SUPPORT
NORTH AMERICA
Customer service representatives are available for questions, problems, or feedback Monday
through Friday, 7:00 am to 5:00 pm Pacific time.
Email:
support.environment@metergroup.com
sales.environment@metergroup.com
Phone:
+1.509.332.5600
Fax:
+1.509.332.5158
Website: