48
SYSTEM
3.2.13.2 REPLACE BATTERY PACK AND COVER
1. Hold the battery pack with the cable as shown in
Figure 27
.
Figure 27 Battery pack with cable oriented for replacement
2. Attach the cable connector to the 2-pin connection inside the battery compartment.
3. Place the battery pack into the battery compartment.
4. Replace the battery cap over the battery pack.
The cap fits very tight around the battery pack. It might need to be angled somewhat to
get it to cover the pack.
NOTE: Be careful to not pinch the cable wire when putting the battery cap on.
5. Replace the two screws in the cap to secure it tight enough for a complete seal.
3.2.14 TEST BUTTON
The
TEST button functions in two ways.
• Reboots the ATMOS 41W and performs a self-test.
• Turns on the Bluetooth module for communication between the instrument and ZENTRA
Utility Mobile.