HYDROS 21

4.  SERVICE

This section contains calibration and maintenance guidelines for the HYDROS 21. 
Troubleshooting solutions and customer support contact information are also provided.

4.1  CALIBRATION

METER factory calibrates the water depth and EC sensors to values stored internally in 
flash memory. The depth sensor is calibrated to known depths of water, and the EC sensor is 
calibrated using potassium chloride (KCl) solutions of known concentrations.

Table 2

 relates EC at 25 °C for various concentrations of KCl, and these values can be used 

to verify the HYDROS 21 EC sensor performance. The value outputs from the HYDROS 21 are 
internally corrected to 25 °C.

Table 2  HYDROS 21 calibration values

Electrical Conductivity 

(mS/cm)

Concentration of KCl in  

Distilled Water 

(g/kg)

100

0.0446

200

0.0930

500

0.2456

1,000

0.5120

2,000

1.0673

5,000

2.8186

10,000

5.8758

20,000

12.2490

4.2  CLEANING

Maintaining the HYDROS 21 is essential for consistent sensor performance. Inspect and 
clean the HYDROS 21 frequently to prevent accumulation of microorganisms, plants, 
sedimentation, or other debris.

4.2.1  PRESSURE TRANSDUCER

The HYDROS 21 ceramic piezoresistive sensor must be protected from sediment and debris. 
If sediment or other debris build up on the pressure transducer, clean the membrane. 

NOTE:  Take care not to damage the pressure sensor.

Follow these steps to clean the HYDROS 21 pressure sensor:
1.  Run water through the side slots of the sensor housing across the ceramic pressure 

transducer to remove debris.

Summary of Contents for HYDROS 21 GEN 1

Page 1: ...HYDROS 21 GEN 1...

Page 2: ...tion 7 3 System 8 3 1 Specifications 8 3 2 Components 11 3 3 Theory 11 3 3 1 Water Depth 11 3 3 2 Temperature 12 3 3 3 Electrical Conductivity 12 4 Service 13 4 1 Calibration 13 4 2 Cleaning 13 4 2 1...

Page 3: ......

Page 4: ...nitoring wells or piezometers surface water monitoring and wetland monitoring The sensor cable is vented to remove the effects of barometric pressure changes The sensor also features a precision therm...

Page 5: ...ns Biofouling can affect sensor performance One way to mitigate the accumulation of microorganisms plants and algae is to wrap the sensor with a copper mesh Inspect and clean the sensor more frequentl...

Page 6: ...em grip to suspend the sensor in its final position Sensor in groundwater well Cap the well Ensure the well is able to vent to atmospheric pressure Surface Water Installation Suspend HYDROS 21 in Wate...

Page 7: ...that these readings are within expected ranges For more instructions on connecting to data loggers refer to Section 2 2 2 2 CONNECTING The HYDROS 21 works seamlessly with METER data loggers The HYDRO...

Page 8: ...te software application to configure the chosen logger port for the HYDROS 21 METER data loggers will automatically recognize HYDROS 21 sensors 3 Set the measurement interval METER data loggers measur...

Page 9: ...ne of the following two options Option 1 1 Clip off the stereo plug connector on the sensor cable 2 Strip and tin the wires 3 Wire it directly into the data logger This option has the advantage of cre...

Page 10: ...MENU button to bring up the Configuration tab NOTE If the PROCHECK does not have this option please upgrade its firmware to the latest version from the METER Legacy Handheld Devices webpage 2 Scroll...

Page 11: ...f 25 C Temperature Range 11 to 49 C Resolution 0 1 C Accuracy 1 C Water Depth Range 0 10 000 mm Resolution 1 mm Accuracy 0 5 of full scale at 20 C NOTE Depth measurement accuracy assumes no abrupt tem...

Page 12: ...le length Cable Diameter 6 mm Connector Types 3 5 mm stereo plug connector or stripped and tinned wires ELECTRICAL AND TIMING CHARACTERISTICS Supply Voltage VCC to GND Minimum 3 6 V Typical NA Maximum...

Page 13: ...1 0 mA Current Drain while asleep Minimum NA Typical 0 3 mA Maximum NA Power Up Time DDI serial Minimum NA Typical 475 ms Maximum 500 ms Power Up Time SDI 12 Minimum 300 ms Typical 475 ms Maximum 500...

Page 14: ...n resin The electronic circuitry is encapsulated in a marine grade epoxy to protect the sensor in corrosive environments Figure 4 HYDROS 21 component diagram 3 3 THEORY This section explains the theor...

Page 15: ...alts in the water and for information about dissolved solids The HYDROS 21 measures EC by applying an alternating electrical current to two electrodes measuring the current flow through those electrod...

Page 16: ...ted to 25 C Table 2 HYDROS 21 calibration values Electrical Conductivity mS cm Concentration of KCl in Distilled Water g kg 100 0 0446 200 0 0930 500 0 2456 1 000 0 5120 2 000 1 0673 5 000 2 8186 10 0...

Page 17: ...roughly with tap or distilled water 4 3 TROUBLESHOOTING Table 3 lists common problems and their solutions If the problem is not listed or these solutions do not solve the issue contact Customer Suppor...

Page 18: ...509 332 5600 Fax 1 509 332 5158 Website metergroup com EUROPE Customer service representatives are available for questions problems or feedback Monday through Friday 8 00 to 17 00 Central European tim...

Page 19: ...16 REFERENCES REFERENCES Sprecher S W 2008 Installing Monitoring Wells in Soils Version 1 0 National Soil Survey Center Natural Resources Conservation Service USDA Lincoln NE...

Page 20: ...customer support 15 E electrical conductivity 8 12 email address 15 F fax number 15 I installation 2 3 groundwater well 2 3 tools required 2 water body 3 P phone number 15 R range 8 references 16 S s...

Page 21: ...Court Pullman WA 99163 T 1 509 332 2756 F 1 509 332 5158 E info metergroup com W metergroup com METER Group AG Mettlacher Stra e 8 81379 M nchen T 49 89 1266520 F 49 89 12665220 E info europe metergr...

Page 22: ......

Reviews: