3

HYDROS 21

Table 1  Installation (continued)

Groundwater Well 
Installation

Suspend HYDROS 21 in Well
Ensure the well has been properly installed, consulting additional resources as 

necessary (e.g., Sprecher, 2008). The well must be 3.5 cm in diameter or larger. 

NOTE:  Well design depends on the desired measurement, such as groundwater flow, 
quality, level, or pressure head. 

Use the Kellem grip to set the desired installation depth. Set the installation 

depth based on the zero point for the pressure transducer.

Lower the sensor down the well and use the hook on the Kellem grip to 

suspend the sensor in its final position. 

Sensor in groundwater well

Cap the well. 

Ensure the well is able to vent to atmospheric pressure.

Surface Water 
Installation

Suspend HYDROS 21 in Water Body
Select a secure location to mount the sensor, such as a bridge abutment or 
pier. If there is not an existing structure, drive a post into the stream bed.

Install sensor protection.

Protect the pressure sensor from noise from water flows. For example, slotted 

well screens protect the sensor from debris and dampens the effect of stream 

current on the pressure measurement.

Suspend sensor from mounting location. 

Sensor suspended in water body

Secure cable, accounting for higher flow conditions to protect sensor cable 

from higher currents and floating debris.

Summary of Contents for HYDROS 21 GEN 1

Page 1: ...HYDROS 21 GEN 1...

Page 2: ...tion 7 3 System 8 3 1 Specifications 8 3 2 Components 11 3 3 Theory 11 3 3 1 Water Depth 11 3 3 2 Temperature 12 3 3 3 Electrical Conductivity 12 4 Service 13 4 1 Calibration 13 4 2 Cleaning 13 4 2 1...

Page 3: ......

Page 4: ...nitoring wells or piezometers surface water monitoring and wetland monitoring The sensor cable is vented to remove the effects of barometric pressure changes The sensor also features a precision therm...

Page 5: ...ns Biofouling can affect sensor performance One way to mitigate the accumulation of microorganisms plants and algae is to wrap the sensor with a copper mesh Inspect and clean the sensor more frequentl...

Page 6: ...em grip to suspend the sensor in its final position Sensor in groundwater well Cap the well Ensure the well is able to vent to atmospheric pressure Surface Water Installation Suspend HYDROS 21 in Wate...

Page 7: ...that these readings are within expected ranges For more instructions on connecting to data loggers refer to Section 2 2 2 2 CONNECTING The HYDROS 21 works seamlessly with METER data loggers The HYDRO...

Page 8: ...te software application to configure the chosen logger port for the HYDROS 21 METER data loggers will automatically recognize HYDROS 21 sensors 3 Set the measurement interval METER data loggers measur...

Page 9: ...ne of the following two options Option 1 1 Clip off the stereo plug connector on the sensor cable 2 Strip and tin the wires 3 Wire it directly into the data logger This option has the advantage of cre...

Page 10: ...MENU button to bring up the Configuration tab NOTE If the PROCHECK does not have this option please upgrade its firmware to the latest version from the METER Legacy Handheld Devices webpage 2 Scroll...

Page 11: ...f 25 C Temperature Range 11 to 49 C Resolution 0 1 C Accuracy 1 C Water Depth Range 0 10 000 mm Resolution 1 mm Accuracy 0 5 of full scale at 20 C NOTE Depth measurement accuracy assumes no abrupt tem...

Page 12: ...le length Cable Diameter 6 mm Connector Types 3 5 mm stereo plug connector or stripped and tinned wires ELECTRICAL AND TIMING CHARACTERISTICS Supply Voltage VCC to GND Minimum 3 6 V Typical NA Maximum...

Page 13: ...1 0 mA Current Drain while asleep Minimum NA Typical 0 3 mA Maximum NA Power Up Time DDI serial Minimum NA Typical 475 ms Maximum 500 ms Power Up Time SDI 12 Minimum 300 ms Typical 475 ms Maximum 500...

Page 14: ...n resin The electronic circuitry is encapsulated in a marine grade epoxy to protect the sensor in corrosive environments Figure 4 HYDROS 21 component diagram 3 3 THEORY This section explains the theor...

Page 15: ...alts in the water and for information about dissolved solids The HYDROS 21 measures EC by applying an alternating electrical current to two electrodes measuring the current flow through those electrod...

Page 16: ...ted to 25 C Table 2 HYDROS 21 calibration values Electrical Conductivity mS cm Concentration of KCl in Distilled Water g kg 100 0 0446 200 0 0930 500 0 2456 1 000 0 5120 2 000 1 0673 5 000 2 8186 10 0...

Page 17: ...roughly with tap or distilled water 4 3 TROUBLESHOOTING Table 3 lists common problems and their solutions If the problem is not listed or these solutions do not solve the issue contact Customer Suppor...

Page 18: ...509 332 5600 Fax 1 509 332 5158 Website metergroup com EUROPE Customer service representatives are available for questions problems or feedback Monday through Friday 8 00 to 17 00 Central European tim...

Page 19: ...16 REFERENCES REFERENCES Sprecher S W 2008 Installing Monitoring Wells in Soils Version 1 0 National Soil Survey Center Natural Resources Conservation Service USDA Lincoln NE...

Page 20: ...customer support 15 E electrical conductivity 8 12 email address 15 F fax number 15 I installation 2 3 groundwater well 2 3 tools required 2 water body 3 P phone number 15 R range 8 references 16 S s...

Page 21: ...Court Pullman WA 99163 T 1 509 332 2756 F 1 509 332 5158 E info metergroup com W metergroup com METER Group AG Mettlacher Stra e 8 81379 M nchen T 49 89 1266520 F 49 89 12665220 E info europe metergr...

Page 22: ......

Reviews: