MD 9070 Insulation/Continuity Digital Multimeter
Specifications
22
o
250 V output voltage: 85 mA.
o
500 V output voltage: 170 mA.
o
1000 V output voltage: 440 mA.
Tester can perform at least 950 insulation tests with
new alkaline batteries at room temperature. These are
standard tests of 1000 V into 1 M
Ω with a duty cycle of
5 seconds on and 25 seconds off.
Low Battery
Approx. 4,6 V.
APO Timing
Idle for 20 minutes.
APO Consumption
50
A typical.
Dimension (LxWxH)
208 x 103 x 64,5 mm with holster.
Weight
635 g with holster.
Accessories
Test probe pair, Alligator clip pair, BRP21S2-C Remote
probe, Holster, User's manual.
Special Features
Record MAX/MIN readings; Display Hold; Backlighted
LCD; VFD V & Hz readings; Lock-Test mode for Insulation
resistance & Earth Continuity Test; BeepJackTM audible &
visible input warning; PI/DAR mode; Compare mode.
ELECTRICAL SPECIFICATIONS
Accuracy is ±(% reading number of digits) or otherwise specified, at
23 °C ± 5 °C
& less than 80% relative humidity.
True RMS voltage & current accuracies are specified from 1 % to 100 % of range or
otherwise specified. Maximum Crest Factor
< 1,70 : 1
at full scale
& < 3,4 : 1
at half scale,
and with frequency components fall within the specified frequency bandwidth for non-
sinusoidal waveforms.
AC Voltage
Range
Accuracy
50 Hz ~ 60 Hz
6,000 V, 60,00 V, 600,0 V, 1000 V
1% + 3d
60 Hz ~ 1 kHz
6,000 V, 60,00 V, 600,0 V, 1000 V
2% + 3d
1 kHz ~ 3 kHz
6,000 V, 60,00 V
2% + 3d
600,0 V, 1000 V
Unspecified
3 kHz ~ 5 kHz
6,000 V, 60,00 V
4% + 5d
600,0 V, 1000 V
Unspecified
Input impedance: 10 MΩ, 110 pF nominal.