MI 2014 Cable Scanner
Section III Cable Scanner operation
17
If one or more faults are detected following
items are displayed:
- the and
signs
- correctly connected pairs
- length if available (depends on the kind of
the fault)
test failed, wires 5,1,2,3,6,7 correctly connected
The related wires, type and distance to fault
(if available) can be viewed by using
and
keys. During displaying faults the sign
is displayed.
test failed, shorted wires 4 and 8 at 2.2m
If the test passed succesfully on the cable
pairs defined in selected Cable Type but
other connections were found that are not in
accordance with the defined cable type
following items are displayed:
- the
and
signs
- the cable length
- correctly connected pairs
test conditionally passed, unselected connected
wires detected
The redundant wire and pairs connections
can be viewed by using
and
keys.
During displaying them the sign is
displayed.
test conditionally passed
(
shorted wires 4 and 8 at
2.2m. The wires are not determined in the selected
cable type)
test conditionally passed
(
Pair 4 is connected. The
pair is not determined in the selected cable type)
Note:
The Fast Test is considered to be performed on cables opened at the far end. If the
cable is terminated by a hub the termination is detected and displayed as a fault.
If a Remote is connected at the far cable end
Remote
is displayed and the test is not
performed.
Fast Test is not applicable when BNC output is selected.
Broken wires (inside the length tolerances) close to the the far end can't be detected
with this test.
Length can't be measured on terminated cables.