Multi LAN 350
Operation
27
Main Single Test Screen
5.3.1 Wire map
Wire Map test verifies the pin to pin wiring and shield continuity. Shield wiring is not
considered in pass/fail decision if UTP cable is selected in Cable Type Menu.
Test procedure:
Make sure that the Remote unit is connected on the other side of the cable.
Select and start the Wire Map test. After the main unit is connected with the Remote
unit and the test is finished, the result is shown on the display together with an overall
Pass/Fail decision. Following actions can now be performed:
ESC key:
return to Single Test main screen
TEST:
repeat the test.
REPEAT:
the Wire Map test is performed continuously.
This enables the operator to find faults like bad contacts or broken wires.
The test can be stopped by using STOP key.
The following table presents different correct and incorrect wirings. Anyl connection fault
can be found with Multi LAN 350. If a fault is found a FAIL warning appears on the
screen.
Condition
Example
Display example
Correct
wiring
Wires and shield are
connected properly.
Open
Wire 1 and shield are open.