Multi LAN 350
Operation
54
Data in received result field
By repeating the steps above locations 1 to 8 can be filled with measurement results.
The following actions can be performed with the entered results:
Data prepared in preset locations
Up, Down, Left, Right
select a result (inverted)
TEST:
clears the selected result
SAVE :
saves the results (refer to chapter 5.11 for more
information)
Note:
Typical causes for a failed tranfer can be:
- distance between Optical power meter and MultiLAN350 too long,
- wrong placing of IR ports of Optical instrument and MultiLAN350,
- obstacles between both IR ports,
- wrong results are sent from the optical instrument (wrong measurement, wrong
wavelength, etc).