Multi LAN 350
Operation
55
5.7 Setting copper cable standard
Refer to chapter 5.13.10 on how to select copper cable measuring mode.
The test standard, cable type and Autotest sequence can be set in the Test Standard
Menu.
There are preprogrammed test parameters for each selected standard or cable type and
Pass/Fail limits. The preprogrammed limits and parameters are used in Autotest and
Single test functions
In the Autotest tests menu the default test sequence can be altered manually by
switching On and Off each test.
For each selected standard, the tests with limits applied are set On by default. Tests
that have no associated limit in the selected cable standard are switched Off by default.
In Appendix 2 a table with preprogrammed test standards and cable types with
associated preprogrammed test and parameter settings can be found. For frequency
dependent limits only data at some typical frequencies are given for the reason of
simplicity. However the complete limit curves are preprogrammed in the instrument.
5.7.1 Test standard selection
After the Standard Menu is selected with the rotary switch the Test Standard main
screen is displayed. In the upper rows the actual set test standard, cable type and NVP
are displayed. The following actions can be performed now:
Up, Down:
select a new test standard
Enter:
enter the Cable Type menu
Test Standard main screen
5.7.2 Cable type selection
Different cable types associated with the selected test standard can be set in this
screen.
In the upper rows the actual test standard, cable type and NVP are displayed. The
following actions can be performed: