Power Network Application Trainer
– Earthing / Grounding network impedance measurement
85
3. Connection is similar to 3 Pole method connection. Only additional connection of ES
(P2) terminal is needed.
- connect the red test probe to ES(P2) terminal and other part to the
pylon
4. Set the parameters by clicking the bottom left dark grey corner
– single mode at 329
Hz
Since exercises on the puzzle uses only for training purposes distance to current (R)
and voltage (r) probes
is not needed. Select “Single test mode “and 329 Hz test
frequency
5. Start the measurement by pressing START key
on the instrument or icon
on the LCD
Expected result: