By observing large differences between maximum ( ), average ( ) and
minimum ( ) recorded values within a single interval we can conclude that a
switching load is present. See figure below.
Figure 4-56: Observing power on switching load
On phase L3 we added a 3
rd
harmonic on current. As a consequence of current
harmonics, nonfundamental apparent power (S
N
) and phase current distortion power
(Di) is introduced as shown in the figure below.
Figure 4-57: Observing power on inductive load with 10% of third harmonics on
current channel I
3
Finish recording by entering the RECORDERS
GENERAL RECORDER menu, and
pressing the F1 (STOP) key. The stored record can be observed on the instrument or
transferred to PowerView. Import your recordings into Metrel PowerView software, as
described in section 4.1.3. All Power and energy related measurements can be
accessed using the
“Power & Energy option” in the “Data explorer” tab. Current, active
consumed power, reactive consumed power and Power Factor are shown below for
example. Note that there is a large difference between Average and Active average
(AvgOn) values in graphs, which clearly indicates that we are dealing with switch load.
Summary of Contents for MI 3399
Page 27: ...Electrical Installation Safety Trainer AD 1 MI 3399 27...
Page 46: ...Electrical Installation Safety Trainer AD 1 MI 3399 46 3 Exit settings...
Page 74: ...Electrical Installation Safety Trainer AD 1 MI 3399 74...
Page 197: ...Power and Voltage Quality Trainer AD 2 MI 3399 197 Figure 4 89 Transient Chart view...
Page 204: ...PAT Appliances Machines Safety Trainer AD 3 MI 3399 204...
Page 249: ...PV Photovoltaic Systems Trainer AD 4 MI 3399 249...
Page 250: ...PV Photovoltaic Systems Trainer AD 4 MI 3399 250 AD 4 MI 3399 PV Photovoltaic Systems Trainer...