Figure 4-82: Inrush Chart view
If we want to observe voltage or current shape, or are interested in phase changes
during such events, then waveform record should be used. Here we can see each
sample of current and voltage. Additionally data is split on 10/12-cycle packets for
which all relevant parameters (RMS U, RMS I, THD, Power) are calculated. Therefore,
the user can check exactly what kind of stress was pushed to the observed object
during such event.