Setup parameters for an ICL-datalogger are shown below:
6.4.1
Example: 25h interval measurement campaign
This example illustrates a measurement campaign, where measurements will be taken with 25 hours
interval. Here, each transmitted measurement will be shifted one hour each day to get measurements
around the clock over a 25 day period. Data will still be transmitted once a day.
498.6
498.62
498.64
498.66
498.68
498.7
498.72
498.74
498.76
498.78
498.8
ER probe thickness [
μ
m], 25h interval measurement campaign
Parameter
Value
Comment
Time for first daily
returned measurement,
HH:MM (UTC)
12:00
Sets time for daily measurement. If the interval between measurements (set in
the row below) is set to be longer than 24h, the time set here will be valid for
the first measurement only.
Time between returned
measurements (hours)
25
Sets interval between measurements to be transferred to WEBservice daily via
satellite. 25 hours are default. 24h intervals or longer intervals are possible, see
section 7.3
Sample interval (min)
60
Cannot be changed by the user. The datalogger performs measurements with 1
hour interval, which are stored in the MasterLink memory.
Eoff timer value (ms)
200
Delay for off-potential measurement. Metricorr recommends 200 ms.
Probe 1
ER-probe
Auto-detected probe by default. It is possible to select probe manually.
Probe 2
ER-probe or
LC sensor
Auto-detected probe by default. It is possible to select probe manually.