Technical Specifications
50,000-count Benchtop Multimeter
49
Technical Specifications
Measuring current
A
DC
current
In "DC" mode, users can measure a direct current value or an alternating
current’s direct component.
A
AC
,
A
AC+DC
TRMS
current
@
Note
In "AC" and "AC+DC" modes, users can measure an alternating current’s rms
value with/without its direct component (no capacitive coupling in "DC" mode).
Results < 0.2 % are not displayed.
@
Note
Specifications for direct current superimposed on a sinusoidal signal in the
40 Hz to 10 kHz band, valid from 10 to 100 % of rate.
For the mA and µA range, the "AUTO PK" mode is always active:
Additional error of 2 % for a peak factor between 1 and 1.5.
Additional error of 15 % for a peak factor between 1.5 and 2.
Peak factor is 5 for a 10 A range.
In 500 mA
DC
and 500 mA
AC+DC
:
Voltage drop
< 600 mV
Protection
600 Vrms per HPC ceramic-type fuse
Fuse
600 V / 10 A / 50 kA Cos
ϕ
> 0.9 (6.3 x 32 mm)
In 10 A and 10 A
AC
(or
AC+DC
):
Voltage drop
< 300 mV
Protection
600 Vrms per HPC ceramic-type fuse
Fuse
600 V / 10 A / 50 kA, Cos
ϕ
> 0.9 (6.3 x 32 mm)
Fuse can be accessed outside unit.
Rate
Precision
Resolution
Input
resistance
Admissible
overload
500 µA
0.2% + 5D
10 nA
about 350
Ω
5 mA
0.2% + 3D
0.1 µA
about 35
Ω
50 mA
0.2% + 3D
1.0 µA
about 5
Ω
500 mA
0.2% + 5D
10.0 µA
about 0.5
Ω
10 A
0.5% + 5D
1 mA
about 0.02
Ω
Rate
Precision
Resolution
Input
resistance
Admissible
overload
500 µA
0.5% + 30D
10 nA
about 350
Ω
5 mA
0.5% + 30D
0.1 µA
about 35
Ω
50 mA
0.5% + 30D
1.0 µA
about 5
Ω
500 mA
0.5% + 30D
10.0 µA
about 0.5
Ω
10 A
2.5% + 30D
1 mA
about 0.02
Ω
Accuracy
40 Hz to 1 kHz
1 to 4 kHz
4 to 10 kHz
500 µA
10 nA
0.5% + 40 D
typ.1%+ 40 D
-
5 mA
0.1 µA
50 mA
1.0 µA
0.5%
1%
2%
500 mA
10 µA
+ 30 D
+ 30 D
+ 30 D
10 A
1 mA
-
-
(cont.)
20 Arms
(30 s max.)
20 Arms
(30 s max.)
Summary of Contents for MTX 3250
Page 2: ......