Technical characteristics of the MX 5006
Benchtop multimeters, 6000 and 60000 points
25
Technical characteristics of the MX 5006 (continued)
CURRENTS
DC current
Particular reference conditions:
µA range: Measuring a strong current for a long time can cause a temperature rise in
some components. In this case, it is necessary to wait some time for the metrological
characteristics specified in µA to be restored.
A
DC
The display indicates "OL" above 19.99A. The
symbol blinks and a beep sounds
above 10A.
(
∗
) Acceptable overload: 10A to 20A for 30s max. with a pause of 5min between 2
measurements. Ambient temp. 35°C max.
Range
Operating
range
Specified
measurement
range
Resolution
Uncertainty
(
±
)
Voltage drop
Protection
6000µA
0 to 6000µA
2 to 6000µA
1µA
0.8% L ± 5 D
25mV/mA
60mA
0 to 60.00mA 0.02 to 60.00mA
0.01mA
0.8% L ± 2 D
3mV/mA
600mA
0 to 600.0mA 0.2 to 600.0mA
0.1mA
0.8% L ± 2 D
0.58mV/mA
6A
0 to 6.000A
0.200 to 6.000A
0.001A
0.8% L ± 3 D
0.05V/A
10A / 20A
∗
0 to 20.00A
0.20 to 20.00A
0.01A
0.8% L ± 2 D
0.05V/A
11A/1000V
fuse
>20kA
AC current
A
AC
RMS
Range
Operating
range
Specified
measurement
range
Resolution
Uncertainty (
±
)
40Hz to 20kHz
(
**
)
Peak factor Voltage drop Protection
6000µA
0 to
6000µA
60 to 6000µA
1µA
1.2% L ±5 D
2.6 to
5mA
25mV /mA
60mA
0 to
60.00mA
6.00 to
60.00mA
0.01mA
2.6 to
50mA
3mV /mA
600mA
0 to
600.0mA
60.0 to
60.0mA
0.1mA
1% L ±3 D
2.6 to
500mA
0.58mV /mA
6 A
0 to
6.000 A
0.600 to
6.000 A
0.001 A
1.2% L ±5 D
2.8 to 5 A
0.05V /mA
10 A / 20 A
∗
0 to
20.00 A
1.00 to
10.00 A
0.01 A
1% L ±3 D
3.7 to 8 A
0.05V /mA
Fuse
11A/1000V
> 20 kA
The display indicates “OL” above 19.99A.
The
symbol blinks and a beep sounds
above 10A
.
Secondary measurements and displays: FREQ (AC coupling) MAX, MIN, PEAK
(
*
) Acceptable overload: 10A to 20A for 30s max. with a pause of 5min between 2
measurements. Ambient temp. 35°C max.
(
**
) Additional uncertainty with the 300Hz filter: see curve on p.29.