Technical characteristics of the MX 5006
Benchtop multimeters, 6000 and 60000 points
27
Technical characteristics of the MX 5006 (continued)
Resistance
Protection: 1414Vpk
Particular reference conditions:
The (+COM) input must not have been overloaded following the accidental application of
a voltage to the input terminals with the switch set to
Ω
or T°.
If this happens, the return to normal may take about ten minutes.
Range
Specified
measurement
range
Resolution
Uncertainty
Measurement
current
Open-circuit
voltage
600
Ω
0 to 600.0
Ω
*
0.1
Ω
0.4% L ±2 D
≈
1mA
6k
Ω
0 to 6.000k
Ω
0.001k
Ω
≈
126.6
μ
A
60k
Ω
0 to 60.00k
Ω
0.01k
Ω
≈
12.6
μ
A
600k
Ω
0 to 600.0k
Ω
0.1k
Ω
0.4% L ±2 D
≈
1.26
μ
A
6M
Ω
0 to 6.000M
Ω
0.001M
Ω
1.5% L ±3 D
≈
240nA
60 M
Ω
0 to 60.00M
Ω
0.01M
Ω
3% L ±3 D
≈
29nA
< 5V
(
*
) REL measurement
Audible
continuity
Protection: 1414Vpk. Response time <100ms
Range
Resolution
Uncertainty
Open-circuit
voltage
Measurement current
600
Ω
0.1
Ω
Audible signal triggered
< 30
Ω
±5
Ω
< 5V
< 1.1mA
Diode Test
Protection: 1414Vpk
Range
Resolution
Uncertainty
Open-circuit
voltage
Measurement current
3V 1mV
Audible signal triggered
< 40mV ±10mV
< 5V
< 1.1mA
Capacitance
Protection: 1414Vpk
Range
Operating range
Specified
measurement
range
Resolution
Intrinsic error
Measurement
current
Measurement
time
6nF
0.100 to
6.000nF
0.100 to
6.000nF
0.001nF
2% L ±15 D
≈
1,26µA
≈
400ms
60nF
0 to 60.00nF
0 to 60.00nF
0.01nF
1% L ±8 D
≈
1,26µA
≈
400ms
600nF 0
to
600.0nF
0 to 600.0nF
0.1nF
1% L ±5 D
≈
1.26µA
≈
400ms
6µF
0 to 6.000µF
0 to 6.000µF
0.001µF
1% L ±5 D
≈
12.6µA
≈
0.125 s/µF
60µF
0 to 60.00µF
0 to 60.00µF
0.01µF
1% L ±5 D
≈
126.6µA
≈
0.125 s/µF
600µF 0
to
600.0µF
0 to 600.0µF
0.1µF
3% L ±5 D
≈
1mA
≈
0.125 s/µF
6mF
0 to 6.000mF
0 to 6.000mF
1µF
4% L ±5 D
≈
1mA
≈
17 s/mF
60mF
0 to 60.00mF
0 to 60.00mF
10µF
6% L ±5 D
≈
1mA
≈
17 s/mF
The use of wires that are very short and shielded is strongly recommended.