Technical characteristics of the MX 5006
Benchtop multimeters, 6000 and 60000 points
29
Technical characteristics of the MX 5006 (continued)
Variation in the
nominal range of use
(*) excluding the 60M
Ω
range
Influence
Quantity of
influence
Range of
influence
Quantity
influenced
typical
MAX
V
DC
mV
0.01% L ±0.2 D / 1°C
0.02% L ±0.25 D /1°C
V
AC
mV, V
LowZ
mV 0.08% L ±0.2 D / 1°C
0.15% L ±0.25 D /1°C
V
DC
0.01% L ±0.1 D / 1°C
0.05% L ±0.1 D / 1°C
V
AC
, V
AC+DC
V
Lowz
0.15% L ±0.1 D / 1°C
A
DC
0.05% L ±0.1 D / 1°C
0.1% L ±0.1 D / 1°C
A
AC
and A
AC+DC
0.08% L ±0.1 D / 1°C
0.12% L ±0.1 D / 1°C
0.01% L ±0.1 D / 1°C
0.1% L / 1°C
Ω
0.05% L / 1°C
0.1% L / 1°C
60M
Ω
0.3% L / 1°C
nF,µF
0.2% L ±0.1 D / 1°C
mF
0.6% L ±0.1 D / 1°C
Hz
0.01% L / 1°C
Temperature
±
2°C + 0.05% L /1°C
Temperature
0°C… 18
28 ... 40°C
Stabilization time
≈
90 min
2 h
Humidity
(without
condensation)
10% ... 80% RH
V
A
Ω
(
*
)
Hz
0
0
1kHz ... 3kHz
4% L
Frequency
3kHz ... 10kHz
V
AC
6% L
Response of the
filter
Filtre 300Hz
-60,00
-50,00
-40,00
-30,00
-20,00
-10,00
0,00
0
50
100
150
200
250
300
350
400
450
Fréquence (Hz)
Er
re
u
r (
%
)
Erreur (%)