Technical characteristics of the MX 5060
Benchtop multimeters, 6000 and 60000 points
31
Technical characteristics of the MX 5060 (continued)
V
AC
RMS
60mV range: Measuring a strong current or measuring a current for a long time may
cause a temperature rise of some components.
Range
Operating
range
Specified
measurement
range
4)
Resolution
Uncertainty
(
±
)
Additional
uncertainty
F(Hz)
1)
Pass band
@ 1kHz
Input
impedance
//< 50 pF
Peak
factor
60mV
2)
0 to
60.000mV
6.000 to
60.000mV
0.001mV
1.5% L
±
35 D
≈
400Hz 10.612M
Ω
3 to
50.0mV
600mV
0 to
600.00mV
60.00 to
600.00mV
0.01mV
1% L
+ 0.25% x
[F(kHz)-1] L
± 30 D
10Hz to
50kHz
(
≈
23%
@100kHz)
10.9M
Ω
3 to
500.0mV
6V
0 to
6.0000V
0.6 to
6.0000V
0.0001V 10.9M
Ω
3 to 5.0V
60V
0 to
60.000V
6.000 to
60.000V
0.001V 10.082M
Ω
3 to
50.0V
600V
0 to
600.00V
60.00 to
600.00V
0.01V 10.008M
Ω
3 to
500.0V
1000V
3)
0 to
1000.0V
60 to
1000.0V
0.1V
0.5% L
+ 0.18% x
[F(kHz)-1] L
± 25 D
45<F<65Hz
0.3% L
typ.
à 100Hz
0.7% L
typ.
à 150Hz
1.8% L
typ.
à 300Hz
30% L
typ.
10Hz to
100kHz
10.008M
Ω
1.42 to
1000.0V
1) See the typical curve of the 300Hz filter on p. 37.
2) This range is accessible only with the
key.
Input impedance: approx. 10.6M
Ω
// 50pF
3) The LCD indicates “+OL” above +1050V, “-OL” below -1050V or above 1050VRMS.
4) From 1kHz, the measurement must exceed 15% of the range.
Secondary measurements and displays: FREQ (AC coupling) MAX, MIN, PEAK
60mV range: Measuring a strong current or measuring a current for a long time may
cause a temperature rise of some components.
AC and DC voltage
AC+DC TRMS
Range
Operating
range
Specified
measuremen
range
4)
Resolution
Additional
uncertainty
DC (
±
)
Uncertainty
AC (
±
)
Additional
uncertainty
F(Hz)
1)
Bandwidth
Input
impedance
//< 50 pF
Peak factor
60mV
2)
0 to
60,000mV
6,000 to
60,000mV
0,001mV
1,5% L
±35 D
≈
400Hz 10,612M
Ω
3 to
50mV
600mV
0 to
600,00mV
60,00 to
600,00mV
0,01mV
0,8% L
+ 0,18% x
[F(kHz)-1]L
±30 D
10Hz to
50kHz
10,9M
Ω
3 to
500mV
6V
0 to
6,0000V
0,6 to
6,0000V
0,0001V
10,9M
Ω
3 to 5V
60V
0 to
60,000V
6,000 to
60,000V
0,001V
10,082M
Ω
3 to
50V
600V
0 to
600,00V
60,00 to
600,00V
0,01V
10,008M
Ω
3 to
500V
1000V
3)
0 to
1000,0V
60 to
1000,0V
0,1V
±
15 D
0,5% L +
0,18% x
[F(kHz)-1]L
±25 D
45<F<65Hz
0,3% L
typ.
at 100Hz
0,7% L
typ.
at 150Hz
1,8% L
typ.
at 300Hz
30% L
typ.
10Hz to
100kHz
10,008M
Ω
1,42 to
1000V
1) See the typical curve of the 300Hz filter on p. 37.
2) This range is accessible only with the
key.
Input impedance: approx. 10.6M
Ω
// 50pF
3) The LCD indicates “+OL” above +1050V, “-OL” below -1050V or above 1050VRMS.
4) From 1kHz, the measurement must exceed 15% of the range
Secondary measurements and displays: FREQ (AC coupling) MAX, MIN, PEAK