Technical characteristics of the MX 5060
34
Benchtop multimeters, 6000 and 60000 points
Technical characteristics of the MX 5060 (continued)
Frequency
Protection: 1414Vpk
Switch set to "Hz",
measurement of the
frequency
of a voltage
Particular reference conditions: 150mV<U<600V
When the switch is set to Hz, the 300Hz filter is not in service
.
Range
Operating range
Specified
measurement range
Resolution
Intrinsic error
60Hz
10.00 to 60.00Hz
10.00 to 60.00Hz
0.01Hz
600Hz
10.0 to 600.0Hz
10.00 to 600.0Hz
0.1Hz
6kHz
0 to 6.000kHz
0.010 to 6.000kHz
0.001kHz
60kHz
0 to 60.00kHz
0.01 to 60.00kHz
0.01kHz
0.1% L ±1 D
Below 10Hz, or if the signal detection level is inadequate, the reading is forced to 0.
*
Measurement of the period in ms can be accessed using the
.
Freq. voltage
or Freq. current
simultaneously,
(secondary display)
Particular reference conditions:
150mV < U < 600V
0.15A
<
I
<
10A
Max. frequency measurable in volts:
60kHz
Max. frequency measurable in amperes: 60kHz
When the switch is set to V
LowZ
, Volts or Ampere, if the 300Hz filter is activated, the
measurable frequency remains within the limits of the PB of the filter.
Below 10Hz, or if the signal detection level is inadequate, the reading is forced: “-----“.
Resistance
Protection: 1414Vpk
Particular reference conditions:
The (+COM) input must not have been overloaded following the accidental application
of a voltage to the input terminals with the switch set to
Ω
or T°.
If this happens, the return to normal may take about ten minutes.
Range
Specified
measurement
range
Resolution
Uncertainty
Measurement
current
Open-circuit
voltage
600
Ω
0 to 600.00
Ω
*
0.01
Ω
0.2% L ±20 D
≈
1mA
6k
Ω
0 to 6.0000k
Ω
0.0001k
Ω
≈
126.6
μ
A
60k
Ω
0 to 60.000k
Ω
0.001k
Ω
≈
12.6
μ
A
600k
Ω
0 to 600.00k
Ω
0.01k
Ω
0.2% L ±20 D
≈
1.26
μ
A
6M
Ω
0 to 6.0000M
Ω
0.0001M
Ω
1.5% L ±30 D
≈
240nA
60M
Ω
0 to 60.000M
Ω
0.001M
Ω
3% L ±30 D
≈
29nA
< 5V
(
*
) REL measurement