Technical characteristics of the MX 5060
Benchtop multimeters, 6000 and 60000 points
35
Technical characteristics of the MX 5060 (continued)
Audible
continuity
Protection: 1414Vpk
Response time <100ms
Range
Resolution
Uncertainty
Open-circuit
voltage
Measurement current
600
Ω
0.01
Ω
Audible signal triggered
< 30
Ω
±5
Ω
< 5V
< 1.1mA
Diode Test
Protection: 1414Vpk
Range
Resolution
Uncertainty
Open-circuit
voltage
Measurement current
3V 0.1mV
Audible signal triggered
< 40mV ±10mV
< 5V
< 1.1mA
Capacitance
Protection: 1414Vpk
Range
Operating range
Specified
measurement
range
Resolution
Intrinsic error
Measurement
current
Measurement
time
6nF
0.100 to 6.000nF 0.100 to 6.000nF
0.001nF
2% L ±15 D
≈
1,26µA
≈
400ms
60nF
0 to 60.00nF
0 to 60.00nF
0.01nF
1% L ±8 D
≈
1,26µA
≈
400ms
600nF 0
to
600.0nF
0 to 600.0nF
0.1nF
1% L ±5 D
≈
1.26µA
≈
400ms
6µF
0 to 6.000µF
0 to 6.000µF
0.001µF
1% L ±5 D
≈
12.6µA
≈
0.125 s/µF
60µF
0 to 60.00µF
0 to 60.00µF
0.01µF
1% L ±5 D
≈
126.6µA
≈
0.125 s/µF
600µF 0
to
600.0µF
0 to 600.0µF
0.1µF
3% L ±5 D
≈
1mA
≈
0.125 s/µF
6mF
0 to 6.000mF
0 to 6.000mF
1µF
4% L ±5 D
≈
1mA
≈
17 s/mF
60mF
0 to 60.00mF
0 to 60.00mF
10µF
6% L ±5 D
≈
1mA
≈
17 s/mF
The use of wires that are very short and shielded is strongly recommended
.
Temperature
Protection: 1414Vpk
Particular reference conditions:
An internal temperature rise may have been caused by:
¾
measurement of a strong current for a long time,
¾
overload of the +COM input with the switch set to T° or
Ω
.
In this case, it is necessary to wait some time to recover the specified metrological
characteristics.
The multimeter must be at the temperature of the room. If not, recovering the
metrological characteristics may take up to 2h. Otherwise, there may be a temperature
offset, because the cold junction temperature reference is a little off.
If there is any doubt, it is possible to check by measuring a known temperature
(
#
example: ambient temperature) with the thermocouple.