Service
6
Benchtop multimeters, 6000 and 60000 points
Service
Maintenance
•
Disconnect everything connected to the instrument and set the switch to
“
OFF
”.
•
Use a soft cloth, moistened with soapy water.
Rinse with a damp cloth and dry rapidly with a dry cloth or forced air.
•
Make sure that no foreign objects interfere with the operation of the device
by which the leads are snapped into place.
Replacing
the fuse
•
Before replacing the fuse, disconnect the instrument from any source of
current.
•
Checking the current fuse:
1. Set the switch to
Ω
.
2. Connect the V socket to the 10A
socket; leave the "COM" socket open.
3. The display unit must indicate a result
< 2
Ω
, if not, replace the fuse.
•
Access to and location of the
fuse
, by opening
the housing:
•
Current protection: fuse, 11A, 1000V > 20kA (10x38)
•
Power supply protection: PTC
Tilt stand
The tilt stand/handle has 2 blue pushbuttons on the sides that are used to
unlock it:
•
Press the pushbuttons simultaneously
•
Adjust the prop to the desired position
•
Release the 2 pushbuttons to lock the handle in position
Communication
interface
The
MX 5060
has a USB communication interface, used:
•
to configure and read the data measured by the instrument (using SX-
DMM software),
•
to recalibrate the instrument.
*
The MX5006 does not have a USB communication port. Only an RS link
is available to calibrate the instrument (after opening it).