Chapter II
44
Benchtop multimeter
6. TECHNICAL
SPECIFICATIONS
Only those values assigned tolerances or limits are guaranteed values. Values without
tolerances are given for information only (French standard NF C 42-670).
{Accuracy : ‘’n% R + nD’’ means ‘’n% of the r n digits’’ as per IEC 485}
6.1. DC
voltages
Selector
switch position
Ranges
Accuracy
Input
impedance
Protection
Resolution
mV
500 mV
0.025%R** + 2D
10 M
Ω
/1G
Ω
*
± 1100 V
PK
***
10 µV
5 V
0.025%R** + 2D
11 M
Ω
± 1100 V
PK
100 µV
V
DC
50 V
0.025%R** + 2D
10 M
Ω
± 1100 V
PK
1 mV
500 V
0.025%R** + 2D
10 M
Ω
± 1100 V
PK
10 mV
1000 V
0.025%R** + 2D
10 M
Ω
± 1100 V
PK
100 mV
* See § 3.3.
*** 1mn max
** At 23 °C
±
2°C
Number of points:
50 000 (or 5 000 - see § 3.3.)
Range selection:
Automatic or manual for the 5 V, 50 V, 500 V,
1000 V ranges
Common mode rejection:
At 50 and 60 Hz, better than 120 dB
Serial mode rejection:
At 50 and 60 Hz, better than 60 dB
Additional error in Pk+/- mode for a pulse of
≥
1 ms: 1% R
±
50 D
For measurements performed on alternative signals, the selected range must tally with the
max. value of the signal peak.
6.2.
AC voltages (AC and AC+DC)
Selector
switch
position
Ranges
Accuracy
Input
impedance
Protectio
n
Resol-
ution
DC*
40 Hz
to 1 kHz
1 kHz
to 4 kHz
4 kHz
to 10 kHz
10 kHz
to 30 kHz
30 kHz
to 50 kHz
5% to 100% of the range
10% to 100% of the range
mV +
SEL
500 mV*
7 % R
+ 30 D
////////////
/////////////
10
M
Ω
/1G
Ω
**
// 100 pF
± 1100
V
PK
***
10 µV
5 V
11 M
Ω
// 100 pF
± 1100
V
PK
100 µV
V
AC
or
50 V
0.3 % R + 30 D
1 % R
+ 30 D
2 % R + 30 D
3 % R
+ 30 D
10 M
Ω
// 100 pF
± 1100
V
PK
1 mV
500 V
10 M
Ω
// 100 pF
± 1100
V
PK
10 mV
V
DC
+
SEL
750 V
/////////////
/////////////
/////////////
/////////////
10 M
Ω
// 100 pF
± 1100
V
PK
100 mV
* AC+DC only.
*** 1 mn max
** See § 3.3.
Number of points:
50 000 (or 5 000 - see § 3.3.)
Range selection:
Automatic or manual for the 5 V, 50 V, 500 V, 750 V ranges
Common mode rejection: At 50 and 60 Hz, better than 80 dB
Additional error according to peak factor:
0.2% for a peak factor of 2 to 3
0.5% for a peak factor of 3 to 6
(Specification given full scale for a squarewave signal pulse 200 µs wide)