Chapter II
Benchtop multimeter
47
6.7.
Diode threshold voltage measurement
Measurable voltages:
0 to 2 V
Measurement current:
1 mA ± 20%
Resolution:
1 mV
Protection:
600 V
RMS
, can be reset automatically
6.8. dB
function
Displays measured values in dBm relative to a resistance reference which can be adjusted
from 1 to 9999 ohms and backed up in non-volatile memory (factory-set to 600 ohms, see
adjustment procedure in § 4.6).
Resolution:
0.01 dB
Absolute error in dB:
0.09 x V
AC
relative error as a percent
Additional computation error:
± 0.01 dB
Measurement range:
10 mV
AC
to 750 V
AC
Protection:
± 1100 V
PK
6.9.
Resistive power function
Displays resistive power relative to a resistance reference which can be adjusted from 1 to
9999 ohms and backed up in non-volatile memory (factory-set to 600 ohms).
Measured function is:
(measured voltage)
2
/ Rref
Resolution:
100 µW
Accuracy:
2 x V
AC
accuracy (%)
Measurement max. voltage :
750 V
AC + DC
Protection:
± 1100 V
PK
Display unit:
VA
6.10. Frequencies
Selector switch setting:
V
AC
, mV, V
DC
, mA, 10 A
Measurement range:
0.62 Hz to 500 kHz
Accuracy:
0.03%
Protection:
1100 V
PK
in V
AC
, mV (1 mn max.), V
DC
modes
600 V
RMS
(F1 + F2) in mA mode
600 V
RMS
(F2) in 10 A mode
Display:
50 000 points
SENSITIVITY
Range
500 mV
5 V to 500 V
500 µA to 500 mA
750 V
10 A
0.62 Hz to 5 kHz*
2% of range
100 V
2 A
5 kHz to 50 kHz
5% of range
250 V
50 kHz to 500 kHz
/////////////
10% of range
/////////////
* rectangular signal
additional positive limit in DC : + 3 % of range, except in 1000 V
DC
range (150 V
additional) and in 10 A
DC
(1.5 A additional)