7.3 Application problems
■■■■■■■■■■■■■■■■■■■■■■
54
■■■■■■■■
Problem
Cause
Remedy
Wrong pH.
In some applications the pH of the measuring
solution can be critical. Check the respective
application documentation for the recommen-
ded pH and measure the pH of the measuring
solution.
Wrong method or incorrect
method parameters.
Check that the method and voltammetric
parameters used for the determination are cor-
rect.
■
Is the method suitable for the intended
concentration?
■
Is potential and time for the potentiostatic
pretreatment correct?
■
Is the expected peak position between
Start potential
and
End potential
of the
sweep?
■
Is the
Potential step
small enough to
have a sufficient resolution of the curve?
Interference from the sam-
ple matrix.
The analyte has to be present in ionic form and
the measuring solution has to be free of inter-
fering substances. For more details see
ter 7.1.2 Measuring solution (page 45)
Electrode is mounted in
reverse position.
Check that the scTRACE Gold electrode
is plugged in with the cor-
rect orientation.
■
The scTRACE Gold has an anti-twist protec-
tion. Nevertheless, it is possible, with a little
force, to mount the electrode incorrectly.
■
The SPE has no anti-twist protection. Make
sure that the electrode surface and the
symbol on the holder point in the same
direction. Only in this position the electrode
has electrical contact in the connection
socket.
Electrode cable is not cor-
rectly assembled.
Check that the electrode cable
is plugged in with the correct
orientation. The cable has an anti-twist protec-
tion. Nevertheless, it is possible, with a little
force, to mount it incorrectly.