Autolab/Electronic load combination
8 |
P a g e
Figure 4 - Configuration of the SMB cables used in combination with the FRA32M module
3
– Autolab part list
The Autolab PGSTAT is used as a voltmeter in combination with the electronic
load. This means that the WE and CE connectors provided by the Autolab must
not be connected to the cell.
The monitor cable is required for the combination with the electronic load and it
must be attached to the front panel of the PGSTAT (see Figure 5).
Dynload
FRA Y
Dynl DSG
FRA V
E-out
FRA X
Warning
Never connect the WE/CE connectors of the Autolab to the electrochemical cell
when working in combination with an electronic load. High currents can
damage the Autolab!
Note
The monitor cable for the PGSTAT204 is not supplied with the instrument and it
must be ordered separately. Please contact
more information.