Autolab/Electronic load combination
19 |
P a g e
8.2
– Programming the load (Kikusui)
Figure 13 shows the overview of the front panel of the Kikusiu PLZ164WA.
Figure 13 – Front panel of the Kikusiu PLZ164WA
Follow these steps to ensure that the load is setup correctly for this application:
1.
Switch on the Kikusui load.
2.
Press SHIFT key and SET/VSET key on the front panel.
3.
Using the rotating knob, highlight the Configuration menu item and press
the ENTER key.
4.
Using the rotating knob, highlight the External menu item and press the
ENTER key.
5.
Using the rotating knob, set change the Control menu item to V and the
LoadOn In item to HIGH.
6.
Press the SHIFT and SET/VSET keys to return to the main menu.
7.
Press the RANGE key to cycle through the different current ranges and set
the active range to the highest available (33 A for the PLZ164WA).
8.
Press the MODE key to cycle through the different operation modes and
make sure that the constant current mode (CC) is selected.
9.
Press the SLEW RATE key and adjust the slew rate to the highest possible
value, using the rotating knob.
10.
Switch the load off and then on again to confirm the changes.