Autolab/Electronic load combination
35 |
P a g e
This window can be used to specify all the settings required to perform an
impedance measurement using the external electronic load.
13.1
– External settings
The parameters on the External section of the FRA editor window are related to
the conversion parameters for setting and measuring the setpoint of the electronic
load as well as the definition of the transfer function.
13.1.1
– Settings for
V
The parameters for the
V signal define the conversion factor between the
output of the FRA2/FRA32M module and the input of the electronic load.
For this application, set the
Unit parameter to A and specify a multiplier (see Figure
Figure 31 – Setting the parameters for the
V output
The value of the multiplier depends on the type of electronic load and the settings
for the Kikusiu PLZ164WA in 33 A current range will be used in this section. For
this load, the multiplier is 0.30303 (10 V / 33 A).
Note
The parameters for other electronic loads are reported in the Appendix of this
document.