Autolab/Electronic load combination
48 |
P a g e
Appendix 1 – Specifications of known electronic loads
The DC and AC conversion settings are specified in Table 2 and Table 3,
respectively.
Load
Current
range (A)
DAC164
1
Conversion slope
ADC164
1
Conversion slope
Kikusui PLZ164WA
33
-3.3
-3.3
Kikusui PLZ664WA
132
-13.2
-13.2
TDI RBL 488
300
-30
-30
TDI WCL 488
1000
-100
-100
Agilent 6060B
60
-6
-6
Agilent 3300
60
-6
-6
Chroma 6300
60
-6
-6
Table 2 – Overview of the DC settings (to be specified in the Hardware setup) for known
electronic loads
Load
Current
range (A)
V Settings
Y Settings
Kikusui PLZ164WA
33
0.303
-3.3
Kikusui PLZ664WA
132
0.0757
-13.2
TDI RBL 488
300
0.0333
-30
TDI WCL 488
1000
0.0100
-100
Agilent 6060B
60
0.166
-6
Agilent 3300
60
0.166
-6
Chroma 6300
60
0.166
-6
Table 3 –Overview of the AC settings (to be specified in the
FRA measurement external
command) for known electronic loads
Note
For information on electronic loads not listed in the Appendix please contact
Metrohm Autolab (