Checking and replacing product parts
■■■■■■■■■■■■■■■■■■■■■■
76
■■■■■■■■
8.4.4
Checking the pump tubing
NOTICE
Interval
The pump tubing must be checked each time before starting work.
Figure 50
Removing the press clamp
1
Press clamp
Removing the press clamp
Prerequisites:
■
The sample robot is disconnected from the power grid.
1
■
Pull the press clamp
on one side outwards by hand.
■
Pull off the other side.
■
Remove the press clamp.
NOTICE
Do not fasten or loosen the hex screws of the press clamps.
Checking the pump tubing
Prerequisites:
■
The sample robot is disconnected from the power grid.