Checking and replacing product parts
■■■■■■■■■■■■■■■■■■■■■■
78
■■■■■■■■
■
Make sure that the press clamp is placed so that it is flush and
positioned cleanly.
8.4.5
Replacing the pump tubing
NOTICE
Use the following pump tubing:
■
Light-colored tubing from Pharm-A-line™ (6.01801.020) with
white press clamp (6.02703.000).
Only pump tubing that has been authorized by Metrohm may be
used.
Dismantling the pump tubing
Prerequisites:
■
The sample robot is disconnected from the power grid.
1 Removing the peristaltic pump tubing
Figure 52
Removing the inlet and outlet tubing
Rotate the tubing counterclockwise and remove it upwards out of
the peristaltic pump.