1-9
9.
Respiration
Normal Physiological Simulation
Baseline Impedances:
500 to 1000 Ohms
Impedance Variations:
0.1, 0.2, 0.5, 1.0, and 3.0 Ohms
Rates:
0 (Apnea), 15, 20, 30, 40, 60, 120 BPM
Output Configuration:
Lead 1, 11, RL-LL
10.
Temperature
30
°
C / 86
°
F, 37
°
C / 98.6
°
F, 40
°
C / 104
°
F
Compatible with YSI 400/700 series.
Accuracy:
±
0.25
°
C
1.3 General Information
Display/Control:
2-digit numeric display keys
5 switches for BP, respiration, temperature and
Power On/Off
ECG Output Connectors
High
Level:
Standard phone jack
Low level:
10 AHA color-coded standard safety banana con-
nectors with detachable banana to snap adapter.
Power:
9 V alkaline battery or battery eliminator
Case:
High impact plastic
Weight:
0.5 kg / 1.1 lbs.
Dimensions:
Height: 47 mm / 1.8 in.
Width: 138 mm / 5.4 in.
Length: 190 mm / 7.5 in.
Standard
Accessories:
PS-416M Patient Simulator
(P.N. 17020)
110 V or 220 V AC Adapter
(P.N. 17021)
Carrying Case
(P.N. 17022)
Snap-to-Banana Adapters (10pk) (P.N. 17023)
User and Service Manual PS-416M (P.N. 17025)
Optional
Accessories:
Unterminated or prewired BP Cable (P.N. 17440)
Unterminated or prewired 400/700
YSI-series Temperature Cable
(P.N. 17443)
Summary of Contents for PS-416M
Page 1: ...1 1 PS 416M User Service Manual PATIENT SIMULATOR...
Page 4: ...1 4 This page intentionally left blank...
Page 6: ...1 6 This page intentionally left blank...
Page 10: ...1 10 This page intentionally left blank...
Page 12: ...2 12 This page intentionally left blank...
Page 20: ...4 20 This page intentionally left blank...
Page 28: ...5 28 This page intentionally left blank...
Page 29: ...5 29 Component Location...
Page 30: ...5 30 Schematic Diagram Part 1...
Page 31: ...5 31 Schematic Diagram Part 2...
Page 32: ...5 32 This page intentionally left blank...
Page 34: ...5 34 This page intentionally left blank...
Page 36: ...5 36...