3 Installation
page
9
ELC N6_N10-V3.1-09/11-GB subject to technical alterations
Mains connection
Fig. 4:
Mains supply
The mains connection of the ELC is
situated at the top right.
The cross section of the earthed wire
must be at least 4 mm².
The three phases must be provided with
an automatic cut-out.
In order to avoid damaging the terminal pins, please do not exceed the
following torque:
•
L1, L2, L3:
0,6 – 0,8 Nm
•
Grounding
conductor:
3,0 – 4,0 Nm