Troubleshooting
Trouble
Possible reason
Action
Insufficient pressure on the connector from the
mushroom nut
Hand tight the connector by
screwing the mushroom nut
O-ring failure
Replace the o-ring
Leaking from Bleed Valve
O-ring failure
Replace the o-ring
Leaking from the adjustment
screw
Ruptured diaphragm
Replace the diaphragm
Ruptured diaphragm
Replace the diaphragm
Insufficient pressure between the diaphragm
and the block
Tighten the 4 mounting
screws to 35lb-in
Adjustment screw not working
Regulating valve spring failure
Spring replacement
Tighten the pressure gauge
NPT connection on the
isolator
Ruptured diaphragm; chemical attacked
pressure indicator connection
Bleeding screw o-ring failure
Ruptured diaphragm
Back Pressure and Pressure Relief Valve
Leaking from the valve block
interface
Pressure Indicator
Leaking fluid from the pressure
indicator NPT connection
Insufficient tightening of the pressure gauge NPT
connection on the isolator
Ship the isolator and
pressure gauge assembly to
our facility
Leaking from the bleeding screw
Module
Leaking from connector