Loading c
onsumables
33
2) Press the left and right edges and ensure that the flow cell is
completely aligned with the flow cell stage.
1
2
2
Figure 12 Loading the flow cell
9. Press the flow cell attachment button A. If the indicator in the flow
cell attachment button illuminates in green, the vacuum pump
works normally. And you will be prompted that the flow cell is
loaded.
After the flow cell is securely seated, do not move it again.
Moving the flow cell might cause the sealing rings to misalign
with holes of the fluidics line.
z
If the flow cell is not attached properly, wipe the flow cell stage
and the back of the flow cell with a clean cloth moistened with 75%
ethanol, and then let it air-dry. Do not press the glass because it might
bring particulate matters and fingerprints or damage the flow cell.
z
If the flow cell accidentally drops to the floor and breaks, handle with
care in case of personal injury.
10. Close all compartment doors.